18
Exporting Data to .csv or .xls Files
•
Using the File Display window, open the fi le you want to export.
•
Click on the tool button.
•
Select a fi le location and enter a name for the fi le.
•
Select a fi le type.
•
Click Save.
A Word About Units
Readings from the T1R Smart Sensor can be displayed in degrees Celsius,or degrees
Fahrenheit, or degrees Kelvin. Select the units you want from the Options | Units men
Field Calibration
Calibration can only be done when there are no sessions stored on the sensor. If there
are any sessions stored on the sensor, upload any data you want and then erase the
session before continuing.
Before leaving the factory, your T1R has been calibrated. However, you may wish to
run a fi eld calibration from time to time.
One-Point Calibration:
-- Computing Calibration Value --
• Place sensor in water. Allow time for sensor to reach thermal equilibrium.
• Using an accurate alternate measuring device, measure the temperature of the
water.
• In
the
Ref deg
box for the fi rst point, enter this temperature.
• Click
fi rst
Measure
button.
• When readings have stabilized to your satisfaction, click the
OK
button in the
pop-up box.
-- Applying Calibration Value --
• Click
the
Apply
button to apply calibration value.
• The computed b value will be transferred to the calibration fi eld.
• Click
OK
to save the value to the sensor.
Two-Point Calibration:
-- First Calibration Point --
• Place sensor in water at fi rst temperature. Allow time for sensor to reach thermal
equilibrium.
• Using an accurate alternate measuring device, measure the temperature of the
water.
• In
the
Ref deg
box for the fi rst point, enter this temperature.