Performing Instrument Tests,
Continued
Temperature Test,
continued
5.
Compare the
Now Temp
value with the desired
Set Temp
value displayed
on the screen. The
Now Temp
value should reach the
Set Temp
value after
a few minutes of starting the test. After reaching the set value, the
Alarm
value, displayed on the screen should turn to
00
.
6.
Press
ESC
to return to the Test Screen. Press
ESC
again to return to the
Main
screen.
Note:
After completing the
Temp
test, the temperature is automatically set to the preset
value of 25ºC. To cool down the heater unit to this preset value, turn
off
the instrument.
The temperature of the heater unit cools down at a rate of about 1ºC per minute. Do not
start a new purification protocol before the preset temperature is reached.
Version Test
To perform the Version Test:
1.
Press
3
from the Test screen.
2.
The Version screen is showing the actual version of the firmware is
displayed.
3.
After confirming the version, press
ESC
to return to the Main screen.
Error Test
To view the error code:
1.
Press
4
from the Test screen.
2.
The Error screen with the error code is displayed.
3.
After viewing the error, press
ESC
to return to the Main screen.