![Invitrogen iPrep Скачать руководство пользователя страница 36](http://html1.mh-extra.com/html/invitrogen/iprep/iprep_user-manual_2085422036.webp)
20
Getting Started,
Continued
Removing the
iPrep
™
Card
Important:
Do not remove the iPrep
™
Card while the instrument is
ON
. This stops the
automated purification protocol and may cause corruption of liquid handling files.
Remove the card as described below.
1.
Ensure the power switch is on the
OFF
position.
2.
Open the iPrep
™
Card Slot.
3.
Push the button located below the card slot to eject the card from the slot.
4.
Pull out the card from the slot and place the card in the box. Store the card
protected from light.
Note:
The card is not completely ejected by pushing the button only. You need to
pull the card out from the slot once the card is ejected slightly from the slot.
Continued on next page