7.8.1 Error codes and diagnostic codes
The drive electronics are capable of rectifying some errors automatically. In this case the status
display will cease to flash. Please switch the remote on and off several times. Wait approx. 5
seconds each time before switching the remote on again. If this does not rectify the error, locate
the error using the flash codes shown below.
FLASH
CODE
FAULT
IMMEDIATE MEASURE
FURTHER HELP
1
Module defective.
-
•
Contact your dealer or
wheelchair provider
2
Lifter raised or
lowered too far (seat
not at driving height)
•
If lifter is raised, lower in
stages until the status
display stops flashing. If
lowered too far, raise lifter
in stages until the status
display stops flashing. If at
all possible, only drive
when the seat is at driving
height.
•
-
Accessory error.
-
•
Contact your dealer or
wheelchair provider.
3
Connection on the left
motor loose/defective
•
Check plug-in
connections.
•
Contact your dealer or
wheelchair provider.
Left motor defective.
•
Check/replace motor
•
Contact your dealer or
wheelchair provider.
4
Connection on the
right motor
loose/defective
•
Check plug-in
connections.
•
Contact your dealer or
wheelchair provider.
Содержание Typhoon II
Страница 1: ...Invacare Typhoon II Electric wheelchair Operating manual Yes You Can...
Страница 64: ...64 8 3 2 Flex and Contour Seats Set the angle by turning the hand wheel 1...
Страница 76: ...76 Loosen the screw 1 using the Allen key Set the desired angle Re tighten the screw...
Страница 117: ...117 Run the seat lifter to the top position Remove legrests Pull out the actuator bolt locking at the belt...
Страница 150: ...150 Order No of this Manual 1521645 DOC Release Date 19 03 08 English...