Troubleshooting

8 Troubleshooting

8.1 Troubleshooting electrical system

Symptom

Possible cause

Remedy

Mains are not connected

Connect mains

Fuse in the control unit is blown

* Replace the control unit

Mains indicator does not light up

Control unit is defective

* Replace the control unit

Motor plug is not fully inserted into the
control unit.

Insert the motor plug properly into the
control unit

The motor is defective.

* Replace the motor

Motor cable is damaged.

* Replace the cable

Mains indicator lights up, but the motor
is not running. The relay in the control
unit makes a clicking noise.

Control unit is defective

* Replace the control unit

Control unit is defective

* Replace the control unit

Mains indicator lights up, but the motor
is not running. No relay sound is heard
from the control unit.

Hand control is defective

* Replace the hand control

Control unit is defective

* Replace the control unit

Control unit is in order except for one
direction on one channel.

Hand control is defective

* Replace the hand control

Motor is running, but the piston rod
does not move.

The motor cannot lift full load.

Motor noise, but no movement of
piston rod.

Piston rod operates inwards and not
outwards.

Motor is damaged

* Replace the motor

* Service and maintenance of the bed must only be performed by personnel who have received the necessary instruction or
training.

Risk of personal injury and damage to the product.

– The bed must be unplugged from the main power source before opening or repairing electrical parts.

WARNING!
Risk of personal injury and damage to the product.

– The bed must be unplugged from the main power source before opening or repairing electrical parts.

1568774-H

21

Содержание SB 910

Страница 1: ...Invacare SB 910 en Bed User Manual This manual MUST be given to the user of the product BEFORE using this product this manual MUST be read and saved for future reference...

Страница 2: ...ation duplication or modification in whole or in part is prohibited without prior written permission from Invacare Trademarks are identified by and All trademarks are owned by or licensed to Invacare...

Страница 3: ...ase of the backrest 13 5 8 1 Quick release backrest CPR optional 13 5 9 Integrated mattress support extension 13 5 10 Lifting pole 14 5 10 1 Positioning the lifting pole 14 5 10 2 Adjusting the handle...

Страница 4: ...ided IMPORTANT Indicates a hazardous situation that could result in damage to property if it is not avoided Tips and Recommendations Gives useful tips recommendations and information for efficient tro...

Страница 5: ...are of persons with small body dimensions WARNING Risk of injury or damage Inappropriate handling of cables incorrect connections and use of unauthorized equipment can cause electrical shock and produ...

Страница 6: ...ng very thick or soft mattresses low density or a combination of these increases the risk A WARNING Risk of falling The user can fall over the edge and get seriously injured if the vertical distance A...

Страница 7: ...ation including technical data HMI XXXXX MD XXXXXXXXXXXXXXXX XXXXXXXXXXXXXXXX XXXXXXXXXXXXXXXX XXXXXXXXXXXXX XXXXXXXXXXXXX XXXXXXXXXXXXX XXXXXXXXXXXXX XXXXXXXXXXXXX xxx kg xxx kg YYYY MM XXXXXX XXXXX_...

Страница 8: ...re listed on the product label and in the technical data section in this manual The bed is not intended for psychiatric patients The bed can be emergency operated by means of the built in accumulator...

Страница 9: ...trical equipment of the bed must not be dismantled or combined with other electrical equipment After each assembly check that all fittings are properly tightened and that all parts have the correct fu...

Страница 10: ...bar B of the side rail away from the locking mechanism CAUTION If the side rail is not properly locked it can fall down Pull push the top bar of the side rail to ensure that the locking system is prop...

Страница 11: ...y to be set 2 Press wanted function button The LED lamp A lights up when the respective function is locked 3 Check that the corresponding function on the hand control is deactivated 4 To unlock a func...

Страница 12: ...nuous connection to the mains Preventive maintenance Batteries must be charged at least every 6 months for maintenance purposes We recommend the batteries to be tested at least once a year Frequent su...

Страница 13: ...e backrest until it is completely down 5 8 1 Quick release backrest CPR optional The CPR function is marked with a label on the bed CPR The quick release handle A is placed under the mattress support...

Страница 14: ...e bed end fitting from the frame This can be done in two steps of respectively 15 and 20 cm 3 Release the centrally located spring and pull out the mattress support extension 4 Tighten both screws aga...

Страница 15: ...ced in one of the four corner tubes 5 11 1 IV Drop rod standard 1 Place the bushing B into the desired corner tube C 2 Position the IV drop rod in the desired height 3 Tighten the finger screw A 5 11...

Страница 16: ...nd disinfectants used must be effective compatible with one another and must protect the materials they are used to clean Never use corrosive fluids alkalines acid etc or abrasive cleaning agents We r...

Страница 17: ...ing IMPORTANT When the bed has been washed before given to a new user it has to be thoroughly inspected Inspection must be done by a trained professional For regular maintenance see Maintenance checkl...

Страница 18: ...k rest function Lock release function working properly 1 7 Height positioning Sections moving correctly motor running properly Run up down in out to outer positions 2 8 Backrest positioning Sections m...

Страница 19: ...hen the hi lo motor has been replaced the mattress support is leaning cannot be set to horizontal Before resetting 1 Disconnect the bed from the mains socket and leave it without power for at least 10...

Страница 20: ...a page 22 Be environmentally responsible and recycle this product through your recycling facility at its end of life Disassemble the product and its components so the different materials can be separa...

Страница 21: ...s not running No relay sound is heard from the control unit Hand control is defective Replace the hand control Control unit is defective Replace the control unit Control unit is in order except for on...

Страница 22: ...ric pressure 800 hPa to 1060 hPa Be aware that when a bed has been stored under low temperatures it must be adjusted to operating conditions before use 9 5 Electrical system Voltage supply Uin 230 VAC...

Страница 23: ...duct uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B Thi...

Страница 24: ...tic site survey should be considered If the measured field strength in the location in which this product is used exceeds the applicable RF compliance level above this product should be observed to ve...

Страница 25: ...800 WLAN 802 11 a n Pulse modulation b 217 Hz 0 2 0 3 9 If necessary to achieve the immunity test level the distance between the transmitting antenna and the Medical Electrical Equipment or System may...

Страница 26: ...Notes...

Страница 27: ...Notes...

Страница 28: ...Fax 353 1 810 7085 ireland invacare com www invacare ie United Kingdom Invacare Limited Pencoed Technology Park Pencoed Bridgend CF35 5AQ Tel 44 0 1656 776 222 Fax 44 0 1656 776 220 uk invacare com w...

Отзывы: