Troubleshooting
Symptoms
Faults
Solution
Patient Lift feels loose.
Mast/Base joint loose.
Tie - Rods are loose.
Refer to ASSEMBLY section in this
manual
Refer to the Base Adjustment in
the CARE AND MAINTENANCE
section in this manual.
Casters/Brakes noisy or stiff
Fluff or debris in bearings.
Refer to Casters in the CARE AND
MAINTENANCE section in this
manual.
Noisy or dry sound from pivots.
Needs lubrication.
Refer to Lubrication in the CARE
AND MAINTENANCE section in
this manual.
Electric actuator fails to lift
when button is pressed.
Hand-control or actuator connector
loose.
Batteries low.
RED emergency stop button pressed
IN.
Battery not connected properly in
to control box.
The connecting terminals are
damaged.
Electric actuator in need of service
or load is too high.
Check connections.
Charge batteries. Refer to
CHARGING THE BATTERIES in the
OPERATION section of this manual.
Rotate RED emergency stop button
CLOCKWISE until it pops out.
Reconnect the battery to the
control box. Refer to CHARGING
BATTERIES in the OPERATION
section of this manual.
Replace the battery pack. Refer to
CHARGING BATTERIES in the
OPERATION section of this manual.
Refer to Electric Actuator
Replacement in the CARE AND
MAINTENANCE section of this
manual. Contact your Dealer.
Unusual noise from Actuator
Actuator is worn or damaged or
spindle is bent.
Refer to ELECTRIC ACTUATOR
REPLACEMENT in the CARE AND
MAINTENANCE section of this
manual Contact your Dealer.
Boom will not lower in
uppermost position.
Boom require a minimum weight
load to lower from the uppermost
position.
Pull down slightly on the lift arms.