Lifter Maintenance

Safety Inspections

A person who is suitably and properly qualified and well acquainted with the design, use and 
care of the hoist should perform periodic inspections.

Inspections should be carried out both routinely (at least as often as specified) and, as necessary.  
All defects and damages that have lead to corrective action should be noted in a log book 
(see page 26) and signed and dated by the inspector. Defects and resulting actions should be 
reported in writing to your supplier.

If inspection reveals that the safety of the hoist is jeopardised in any way, the Owner should 
be notified immediately, the hoist should be taken out of service and not used again until the 
problem has been rectified. Contact the supplier to arrange servicing.

A Guide to Maintenance Procedures

A regular visual check of the hoist and attachments is essential. Invacare Australia Pty Ltd 
permits servicing only by an Authorised Service Agent in accordance with the following schedule 
of Service Types.

Item

Pre-Delivery

Daily

Minor Service 

(After every 

3 Months of 

Service)

Major Service 

(After every 

6 Months of 

Service)

PRE-DELIVERY

Confirm all components are included

X

Inspect parts for possible damage in transit

X

COMPLETE LIFTER

General inspection of Lifter.  Visually inspect the patient lifter for 
external damage or wear. Scratches and chips.

X

X

X

X

Retard rusting by applying spray paint (Dulux ‘Special White’ Code: 
81880)

X

X

BASE AND CASTORS

Inspect for missing hardware

X

X

X

Base opens/closes with ease.

X

X

X

Inspect castors and axle bolts for tightness. Make sure tightly attached 
to legs.

X

X

X

Inspect castors for smooth swivel and roll

X

X

X

Check castor brakes operate efficiently

X

X

X

Overhaul castors and repack with grease

X

X

Check tightness of bolt in leg bushes

X

Check linkages are secure in leg adjuster ball joints

X

MAST

Mast MUST be securely assembled to boom

X

X

X

Inspect for bends or deflections.

X

X

X

Check sling hooks for wear or deflection.

X

X

X

X

ELECTRIC ACTUATOR ASSEMBLY

Check hand control and cycle to ensure smooth quiet operation. Lift 
adjustment on all Lifters. Width also for those with electric leg adjustment.

X

X

X

X

Emergency lowering function (electric and mechanical)

X

X

X

X

Check for wear or deterioration. (IF DAMAGED, RETURN TO FACTORY)

X

X

X

Inspect hardware on mast and boom.

X

X

X

Remove dust and dirt from actuator with cloth

X

X

Содержание RPS 440ee

Страница 1: ...R This manual MUST be given to the user of the patient lift USER BEFORE using this patient lift read this manual and save for future reference SERVICE AGENT Only Invacare authorised service agents are...

Страница 2: ......

Страница 3: ...Emergency Stop Button OFF while transmission is in progress This product meets EN60601 1 2 Collateral Standard Electromagnetic Compatibility WARNING notices as used in this manual apply to hazards or...

Страница 4: ...ic RPS 440ee User s Manual Special Notes ALWAYS be aware of the Lift Arms Injury to the patient and or assistant may occur ALWAYS be aware of the Footrest especially the patient s position on the foot...

Страница 5: ...ocking Unlocking the Rear Swivel Casters 12 Raising Lowering Stand Up Lift 12 Closing Opening the Legs of the Base 12 Mounting Battery Charger 13 Charging the Batteries 14 Transferring From 15 Lifting...

Страница 6: ...reach at min height of C S P NA 11 Max hoisting reach NA 12 Base Base Radial type 1 Max external length 1140 2 Max internal length 820 3 Min internal width 575 4 Min external width 675 5 Max internal...

Страница 7: ...acare Stand Up Lift Slings and accessories designed by other manufacturers or other Invacare slings are not to be utilized as a component of Invacare s stand up lift system Use of these products is pr...

Страница 8: ...he sling is properly connected to the attachment points of the stand up lift If any attachments are NOT properly in place lower the patient back onto the surface and correct this problem During transf...

Страница 9: ...n the European Market It indicates that the product complies with applicable European Directives related to health safety environment and consumer protection Safe Working Load Represents the maximum l...

Страница 10: ...es you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific RPS 440ee User s Manual Documentation With each lifter you should receive the following documents this Instruction Manual a...

Страница 11: ...using the previously removed bolts Note Ensure these bolts are firmly tightened Attach the rod ends as shown in 2 Cut cable tie securing actuator 3 and swing actuator forward so that it is out of the...

Страница 12: ...clevis and restrict the pivot Attaching the kneepad Attach the kneepad to the kneepad receiver by sliding the two posts into the tube retainers of the kneepad receiver assembly and at the same time t...

Страница 13: ...ion Invacare is a registered trademark andYes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific Ensure control box emergency push button is set to the off position out If not t...

Страница 14: ...patient NOTE If the Stand Up Lift is in the full UP position it may be necessary to pull down slightly on the lift arms before the mast will lower EMERGENCY STOP Press the RED button on the control b...

Страница 15: ...h mounting bracket on the wall at the desired position 2 With a pencil mark the MIDDLE hole position 3 Measure down 6 1 2 inches from the pencil mark and drill one 1 mounting hole 4 Install the BOTTOM...

Страница 16: ...ay from the control box 3 Place the battery on the battery charger as shown in the following diagram Make sure there is an audible click NOTE The charge LED will illuminate When charging is complete c...

Страница 17: ...elf when the patient is initially lifted from and transferred to a chair bed or any stationary object Transferring From Lifting Preparation NOTE The stand up lift may be operated by one 1 healthcare p...

Страница 18: ...s MUST be in the locked position BEFORE lowering the patient into the wheelchair STANDING SLINGS DO NOT use the standing sling in combination with the stand up lift as a transport device It is intende...

Страница 19: ...his her own weight However since medical conditions vary Invacare recommends that the healthcare professional evaluate the need for assistance and determine whether more than one 1 assistant is approp...

Страница 20: ...button to raise the patient above the surface of the bed The patient should be elevated just high enough to clear the bed with their weight fully supported by the lift 2 Press the DOWN button and low...

Страница 21: ...lings Unhook the transport sling from the bottom attachment points on the stand up lift Lift up on the patient s legs and remove the thigh supports from underneath the patient If desired unhook the tr...

Страница 22: ...n accordance with the following schedule of Service Types Item Pre Delivery Daily Minor Service After every 3 Months of Service Major Service After every 6 Months of Service PRE DELIVERY Confirm all c...

Страница 23: ...D HARDWARE CHECK ALL SLING ATTACHMENTS each time it is used to ensure proper connection and patient safety X X X X Inspect sling material for wear X X X X Inspect straps for wear X X X X CLEANING When...

Страница 24: ...to lift when button is pressed Hand control or actuator connector loose Batteries low RED emergency stop button pressed IN Battery not connected properly in to control box The connecting terminals are...

Страница 25: ...lts for tightness Remove all debris etc from the wheel and swivel bearings If any parts are worn replace these parts immediately If you question the safety of any part of the lift contact your Dealer...

Страница 26: ...o clean the patient lift The lift can be cleaned with non abrasive cleaners Electric Actuator Replacement 1 Remove the nut washer and shoulder bolt that secure the electric actuator to the mast mounti...

Страница 27: ...efective unit or parts MUST be returned for warranty inspection using the serial number when applicable as identification within 30 days of return authorization date Do not return products to our fact...

Страница 28: ...f inspection reveals that user safety is jeopardised in any way from use of the patient lifter Ensure ongoing servicing after the Warranty Period is maintained by agreement between you and an Authoris...

Страница 29: ...onTaken Inspected By Signature Date ServiceType Pre Delivery Minor Periodic Inspection Major RPS 440ee User s Manual All information is subject to change without prior notice Photographs are for illus...

Страница 30: ...y Signature Date ServiceType Pre Delivery Minor Periodic Inspection Major All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from pro...

Страница 31: ...y Signature Date ServiceType Pre Delivery Minor Periodic Inspection Major RPS 440ee User s Manual All information is subject to change without prior notice Photographs are for illustration purposes on...

Страница 32: ...y Signature Date ServiceType Pre Delivery Minor Periodic Inspection Major All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from pro...

Страница 33: ...y Signature Date ServiceType Pre Delivery Minor Periodic Inspection Major RPS 440ee User s Manual All information is subject to change without prior notice Photographs are for illustration purposes on...

Страница 34: ...y Signature Date ServiceType Pre Delivery Minor Periodic Inspection Major All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from pro...

Страница 35: ...subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark andYes you can is a trademark of Invacare...

Страница 36: ...ax 0508 807 788 Email sales invacare co nz Web www invacare co nz InvacareAsia Ltd 1 Lenton Place PO Box 5002 North Rocks NSW 2151 Australia Phone 61 2 8839 5330 Fax 61 2 8839 5331 Email asiasales inv...