Troubleshooting
7.2
Troubleshooting: Mechanical
SYMPTOM
PROBABLE CAUSE
DEALER SOLUTION
USER SOLUTION
Bed does not stay in place.
Locking casters not locked.
Lock casters. Refer to 4.5
Installing Casters, page 21.
Lock casters.
Bed sections do not move.
Pull tube.
Ensure pull tubes are properly
connected. Check pull tube(s) for
binding. Call dealer for repairs.
Check pull tube (s) for binding.
Call dealer for repairs.
Bed ends move in opposite
directions (electrical or manual
bed).
Drive shaft connected to wrong
gear box shaft.
Reinstall drive shaft. Refer to 4.8
Assembling/Installing/Removing
Drive Shaft, page 25.
Call dealer.
Hi/Lo function does not work.
Drive shaft disconnected.
Reinstall drive shaft. Refer to 4.8
Assembling/Installing/Removing
Drive Shaft, page 25
Call dealer.
Pressing function on pendant
results in motion other than that
intended.
Motor cable connections to
junction box incorrect.
Verify connections.
Call dealer.
1123842-I-05
39