Serial Number
Indoor Use ONLY
DO NOT use oil or grease
Keep dry
-29C
MIN
70C
MAX
Transport and Storage Temperature
Input Fitting
Output Fitting
This product complies with Directive
93/42/EEC concerning medical devices.
The launch date of this product is stated in
the CE declaration of conformity.
1.2
Intended Use
The Precise R
x
™ Pediatric Flowmeter is intended to provide
supplemental oxygen to patients with low-flow prescriptions
in a non-acute setting such as the home or long-term care
facility. The device is not intended to be life supporting or
life sustaining.
WARNING!
Risk of Injury or Damage
Use of this product outside of the intended use
and specifications has not been tested and may
lead to product damage, loss of product function,
or personal injury.
– DO NOT use this product in any way other than
described in the specifications and intended
use sections of this manual.
1.3
Indications for Use
To provide supplemental oxygen to patients with low-flow
prescriptions in a non-acute setting.
1.4
Contraindications
There are no known contraindications.
1.5
Technical Description
The device is a free standing, non-electrically powered,
accessory for connection to the output fitting of an Invacare
5 liter output oxygen concentrator that delivers low flow
rates of less than one (1) liter per minute to a pediatric
patient through a pediatric nasal cannula.
Input oxygen gas at an input pressure of 5.0 +/- 10% PSI
and at a minimum flow rate of 2 liters per minute by a
compatible Invacare 5 liter output oxygen concentrator is
delivered to the device input through an external extension
tubing accessory. The input gas is routed to an internal
bleed orifice for maintaining a minimum flow output load
to the oxygen concentrator for maintaining adequate
oxygen concentration values of the delivered gas by the
1150712-D
5