Invacare® PreciseRx™ Pediatric Flowmeter
3.4
Setup With a Humidifier Bottle
A
Pediatric Flowmeter
B
Input Fitting of Pediatric Flowmeter
C
Output Fitting of Concentrator
D
Concentrator
E
2.1 m (7 ft) Oxygen Tubing from Concentrator to
Flowmeter
F
Output Fitting of Flowmeter
G
2.1 m (7 ft) Oxygen Tubing from Flowmeter to
Concentrator
H
Humidifier Bottle
I
Inlet of Humidifier Bottle
1.
Connect a 2.1 m (7 ft) maximum oxygen tubing
E
from
the output fitting
C
of the concentrator
D
to the input
fitting
B
of the Pediatric Flowmeter
A
.
2.
Follow the directions supplied by the humidifier
bottle
H
manufacturer for instructions on filling and
connecting to oxygen tubing.
3.
After filling the humidifier bottle with water, place
it in the humidifier compartment on the Invacare
concentrator
D
(see concentrator user manual for
mounting the bottle).
4.
Connect 2.1 m (7 ft) maximum oxygen tubing
G
from
the Pediatric Flowmeter
A
output fitting
F
to the
inlet of the humidifier bottle
I
(using an appropriate
adapter if applicable).
5.
Connect a 7.6 m (25 ft) maximum oxygen tubing to the
outlet of the humidifier bottle (not shown).
6.
Attach the opposite end of the oxygen tubing to a
pediatric nasal cannula using an appropriate oxygen
tubing connector (not shown).
CAUTION!
– DO NOT coil crush resistant oxygen tubing on
COLD floor or condensation will form inside
the tubing.
14
1150712-D