SECTION 12—WIRING ASSEMBLIES
81
SECTION 12—WIRING ASSEMBLIES
Replacing Wiring Assemblies
DANGER
To prevent electrical shock, ALWAYS disconnect compressor from electrical
outlet before servicing.
Full View of Wiring Harnesses
NOTE: For this procedure, refer to FIGURE 12.1.
FIGURE 12.1
Replacing Transformer Assembly
NOTE: For this procedure, refer to FIGURE 12.2 on page 82.
Tools required for this procedure are:
• Phillips screwdriver
1. Remove the cabinet. Refer to Removing/Installing the Cabinet on page 25.
2. Remove the control panel. Refer to Removing/Installing Control Panel on page 29.
3. Remove the two mounting screws that secure the transformer to the base (Detail “A”).
4. Disconnect the connector from the circuit board (Detail “B”).
5. Remove the transformer assembly wire harness from the HomeFill compressor.
HF Wire
Assembly Harness
Bleed Resistor
Assembly 1
Transformer
Assembly Wire
Harness
Bleed Resistor
Assembly 2
NOTE: The bleed
resistors in this view are
for units made before 7/
3/06. On units made
after 7/3/06, bleed
resistors are on the PC
board with a separate