SECTION 3—CHECKLIST AND CYLINDER INSPECTION
19
Hydrostatic Testing Date
Aluminum cylinders MUST undergo testing every ten years.
WARNING
Risk of Injury
To reduce the risk of injury:
- DO NOT fill cylinders that have not been tested in the past ten years. Otherwise,
serious personal injury may result. Contact your dealer/Home Care equipment
provider for replacement.
External Examination
1. Examine the outside of the cylinder for the following conditions, and replace the
cylinder if they exist:
• Dents or dings
• Arc burns
• Oil or grease
• Any other signs of damage that might cause a cylinder to be unacceptable or
unsafe for use.
2. Examine the cylinder for evidence of fire or thermal damage. Evidence includes
charring or blistering of the paint, or other protective coating or heat sensitive
indicator. If fire or thermal damage is found, replace the cylinder.
3. Inspect the Invacare valve for the following and replace if found:
• Debris, oil or grease
• Noticeable signs of damage
• Signs of corrosion inside the valve
• Signs of excessive heat or fire damage