Service
1.
Turn key to switch mobility device off.
2.
Press and hold keys.
3.
Turn key to switch mobility device on.
Mobility device enters setup mode after two seconds.
Enter next display with
key.
Enter next display with
key.
Enter next display with
key.
Setting counter
This chapter applies for mobility devices with LCD
display.
1.
Enter setup settings as described in
Entering setup (LCD
Display), page 16
.
Default setting is 0. You can set a km value as
service interval.
If counter value is greater than set value,
symbol will flash for one minute, when key is switched
on.
+
0 (OFF)+ 500 + 1000
+1500 + 2000 ..........
+ 9500 + 9999 (max)
-
9999 – 9500 – 9000 –
8500 – 8000 ......... –
500 – 0 (OFF)
2.
Press and hold key for 2 sec. to erase counter.
3.
Press key to increase value.
4.
Press key to decrease value.
5.
Press key to save and enter next page.
Setting battery gauge
This chapter applies for mobility devices with LCD
display.
Setting battery gauge is mandatory when retrofitting
operating console.
Adjust the parameter in battery gauge to suit different types
and sizes of batteries.
AGM battery
Gel battery
(1) big size
batteries (>
50 A)
(2) small
size batteries
(< 50 A,
default)
(3) small size
batteries (<
50 A)
(4) big size
batteries (>
50 A)
1.
Press key to increase value.
2.
Press key to decrease value.
3.
Press key to save and enter next page.
Setting wheel diameter
This chapter applies to mobility devices with LCD
display.
Setting wheel diameter is mandatory when
retrofitting operating console.
Set value according to wheel diameter 11”, 12”, 13”, 14” or
15” for correct display of speed.
1603798-B
17