Comet

Replacing lifter

1.

Remove batteries.

2.

Disconnect lifter cable at wiring harness connecting plug
in motor enclosure.

3.

Remove screws (1) on lifter (2).

4.

Replace lifter.

5.

Install parts in reverse order.

6.

Test all functions.

Replacing lifter controls

1.

Remove both screws (1) from lifter controls (2).

2.

Replace lifter controls (2).

3.

Install parts in reverse order.

4.

Test all functions.

Replacing fuses in lifter controls

1.

Loosen and remove screws on lifter controls shroud.

2.

Replace fuse (1).

3.

Install parts in reverse order.

4.

Test all functions.

4.5.5 Replacing power module

CAUTION!
Risk of accident

When parking mobility device on antitippers it is
no longer slowed by motor brake. Mobility device
can roll away out of control.

– Place mobility devices’s rear frame on a

supporting wooden block before you remove
wheels.

CAUTION!
Risk of burning if power cable is shorted

– Turn off power completely before removing any

power supply components of mobility device.
Therefore take out batteries.

– Avoid bridging of contacts during measurements

on live electrical components.

CAUTION!
Any changes to the drive program can affect the
driving characteristics and the tipping stability
of the mobility device.

– Changes to drive program may only be carried

out by trained Invacare® specialist providers.

– Invacare® can only give a warranty for safe

mobility device driving behaviour - especially
tipping stability - for unaltered standard drive
programs.

CAUTION!
Risk of accident. Accidental rolling can lead to
accidents.

– Secure mobility device against rolling away.

The controls system is supplied with a standard drive
program. If you have carried out customer-specific
modifications to the drive program, you will have to
make these changes again after installing the new
power module.

When disassembling, note position of small parts
such as screws and washers. Put small parts down
so that they can be reassembled in right sequence.

The plugs on power module cannot be wrongly
connected because all plugs have a different size and
only fit in one socket.

10 mm wrench

17 mm wrench

Rubber hammer

Phillips screwdriver

Supporting wooden block

To adapt the drive program: programming
software or hand programming device and
controls system installation manual, available
from Invacare.

1.

Secure mobility device against rolling away.

2.

Place supporting wooden block under frames at battery
holder height.

3.

Remove seat.

4.

Remove shroud.

14

1603798-B

Содержание Comet ALPINE PLUS

Страница 1: ...Invacare Comet Series CometPRO CometALPINE CometULTRA en Scooter Service Manual DEALER Keep this manual The procedures in this manual MUST be performed by a qualified technician...

Страница 2: ...e or in part is prohibited without prior written permission from Invacare Trademarks are identified by and All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless ot...

Страница 3: ...hroud 18 4 6 2 Replacing front shroud 18 4 7 Lighting unit 19 4 7 1 Replacing headlight 19 4 7 2 Replacing front direction indicators 19 4 7 3 Replacing center brake light 20 4 7 4 Replacing rear ligh...

Страница 4: ...y executed maintenance or overhaul work lead to the exclusion of all liability on the side of INVACARE If you have any problems or questions contact Invacare Service 1 2 Notes on shipping If the mobil...

Страница 5: ...very heavy This results in injury hazards to your hands Note the high weight of some components This applies especially to the removal of drive units batteries and the seat CAUTION Injury hazard if th...

Страница 6: ...for safe vehicle driving behavior especially tipping stability for this standard drive program Mark all current settings for the mobility device seat armrests backrest etc and the associated cable con...

Страница 7: ...station Manual cleaning Surface of used device Before repair or reconditioning Use saturated towel to apply cleaning detergent and remove residues after impact Cleaning and disinfection Disinfection...

Страница 8: ...Check main fuse See user manual for main fuse position Check cable between modules for loose connections or damage See Replacing power module Status display on operating console does not illuminate Op...

Страница 9: ...here is a defect in the braking coil or in the cabling Check the magnetic brake and cabling for open or short circuited circuitry Contact your Invacare provider 6 No neutral position when switching sc...

Страница 10: ...Check front wheels for tight fit float and side play Adjust replace Check pneumatic tire Repair or replace if damaged See user manual Tighten screws Wheel suspension and wheels Check tracking and stee...

Страница 11: ...urns if power cable is shorted Turn off the power completely before removing any power supply components of mobility device Therefore take out batteries Avoid bridging of contacts during measurements...

Страница 12: ...rrectly connected 1 Unplug battery connector 2 Remove operating console as described in chapter 4 5 1 Replacing operating console page 11 3 Replace circuit board 4 Make sure all pins are correctly con...

Страница 13: ...operating console 12 Connect potentiometer A and drive lever mounting B Tighten screw C 13 Install parts in reverse order 14 Test all functions trial run 4 5 4 Replacing lifter lifter controls CAUTION...

Страница 14: ...affect the driving characteristics and the tipping stability of the mobility device Changes to drive program may only be carried out by trained Invacare specialist providers Invacare can only give a w...

Страница 15: ...fter 30 seconds H Lighting I Low speed mode J Drive lever Status display The ON OFF diode 1 is used as a fault display status display It will flash if there is a problem with the scooter The number of...

Страница 16: ...indication 1If this symbol starts flashing for one minute every time the scooter is switched on contact your provider Battery indication gauge Low battery indication Every time the scooter is activate...

Страница 17: ...for 2 sec to erase counter 3 Press key to increase value 4 Press key to decrease value 5 Press key to save and enter next page Setting battery gauge This chapter applies for mobility devices with LCD...

Страница 18: ...ng console Adjust the parameter in battery gauge to suit different types and sizes of batteries AGM battery Gel battery 1 big size batteries 50 A 2 small size batteries 50 A default 3 small size batte...

Страница 19: ...sconnect battery cable A 4 Remove rear steering column shroud See 4 8 4 Replacing steering column page 23 5 Remove screw A on headlight under front shroud 6 Disconnect direction indicator cable B at w...

Страница 20: ...t they can be reassembled in the right sequence Phillips screwdriver 1 Remove seat 2 Remove shroud 3 Disconnect battery cable A 4 Disconnect plug 4 5 Remove bolts A on rear shroud 6 Lift rear shroud u...

Страница 21: ...unction 4 8 Wheels 4 8 1 Replacing wheel suspension CAUTION Risk of accident Accidental rolling can lead to accidents Secure scooter against rolling away When disassembling note the position of small...

Страница 22: ...ft hand shock absorber If the right hand shock absorber is to be replaced you do not need to remove the drive motor 10 mm wrench 17 mm wrench Phillips screwdriver 5 mm Allen keys 2x Supporting wooden...

Страница 23: ...cing steering column CAUTION Risk of accident Accidental rolling can lead to accidents Secure mobility device against rolling away When disassembling note the position of small parts such as screws an...

Страница 24: ...9 1 Replacing hand brakes page 26 14 Loosen and remove screw at clamp 1 15 Loosen and remove fixing bolt for steering column 16 Pull steering column out upwards 17 Install parts in reverse order Tight...

Страница 25: ...en three screws D E that fix two steering links 10 Replace suspension 11 Install parts in reverse order 4 8 6 Replacing and Adjusting Central Rear Shock Absorber 17 mm wrench 15 mm wrench Wooden block...

Страница 26: ...d Always let all the air out of the tire before removing the rim 3 Loosen and remove 4 screws 1 4 Remove both wheel rim halves out of tire and remove inner tube 5 Repair inner tube and refit in wheel...

Страница 27: ...ake cable CAUTION Risk of accident Accidental rolling can lead to accidents Secure mobility device against rolling away When disassembling note position of small parts such as screws and washers Put s...

Страница 28: ...ing link up by 0 5 cm Wheels stand turned inwards by 1 C from neutral position B 3 Install parts of steering link and or coupling rod in reverse order 4 Test all functions trial run 4 9 4 Replacing cu...

Страница 29: ...CAUTION Risk of accident When parking mobility device on antitippers it is no longer slowed by motor brake Mobility device can roll away out of control Place mobility device rear frame on a supporting...

Страница 30: ...15 Remove disengaging lever 2 16 Place supporting wooden block under swing arm Replacing drive motor 1 Loosen three Allen screws A 2 Lift motor out of holder and take it out of the frames to the side...

Страница 31: ...brushes 1 Depending on condition of brush and spring either install brushes in exactly the same position from which they were taken or fit new brushes 2 Install plastic caps and tighten them firmly 3...

Страница 32: ...ve shroud See Removing shroud 3 Remove securing bolt nut and washers of seat support tube 4 Replace seat support tube 5 Adjust seat height 6 Reinsert securing bolt nut and washers and tighten 4 12 Acc...

Страница 33: ...Notes...

Страница 34: ...Notes...

Страница 35: ...Notes...

Страница 36: ...ield Place Mt Wellington 1060 New Zealand Phone 0800 468 222 Fax 0800 807 788 sales invacare co nz www invacare co nz United Kingdom Invacare Limited Pencoed Technology Park Pencoed Bridgend CF35 5AQ...

Отзывы: