Azalea®
6
Troubleshooting
6.1
Troubleshooting electrical system
Symptom
Possible cause
Remedy
Mains are not connected
Connect mains
Fuse in the control unit is blown
Replace the control unit
Mains indicator does not light up
Control unit is defective
Replace the control unit
Motor plug is not fully inserted into the
control unit.
Insert the motor plug properly into the
control unit
The motor is defective.
Replace the motor
Motor cable is damaged.
Replace the cable
Mains indicator lights up, but the motor is not
running. The relay in the control unit makes a
clicking noise.
Control unit is defective
Replace the control unit
Control unit is defective
Replace the control unit
Mains indicator lights up, but the motor is not
running. No relay sound is heard from the
control unit.
Hand control is defective
Replace the hand control
Control unit is defective
Replace the control unit
Control unit is in order except for one
direction on one channel.
Hand control is defective
Replace the hand control
Motor is running, but the piston rod does not
move.
The motor cannot lift full load.
Motor noise, but no movement of piston rod.
Piston rod operates inwards and not
outwards.
Motor is damaged
Replace the motor
Risk of personal injury and damage to the product.
–
The wheelchair must be unplugged from the main power source before opening or repairing electrical parts.
WARNING!
Risk of personal injury and damage to the product.
–
The wheelchair must be unplugged from the main power source before opening or repairing electrical parts.
32
1580313-B
Содержание Azalea Assist
Страница 36: ...Azalea 7 7 3 Rea Azalea equipped with Tall kit 4 5 4 150 3 2 1 50 24 4 2 200 50 16 5 3 200 36 1580313 B...
Страница 37: ...Notes...
Страница 38: ...Notes...
Страница 39: ...Notes...