4 Repair
Repairs require a high level of expertise. These assembly
instructions therefore break down the various tasks into
3 categories:
Requirement
Symbol
Easy – technical understanding required
●○○
Medium – technical knowledge required
●●○
Difficult – technical knowledge and expertise
in assembling bath lifts required
●●●
The required tools and their sizes are listed before the
instructions.
Tool
Symbol
Phillips head screwdriver
Slot head screwdriver
Torx screwdriver
Snap ring pliers
WARNING!
– To avoid the risk of contamination, clean and
disinfect the bath lift before carrying out
repairs.
IMPORTANT!
– Repairs may only be carried out by an
authorized dealer.
– If possible, continue to use the old
identification label; if this is not possible,
the new identification label must contain
the same information and the old serial
number. (Replacement of spare parts with
serial numbers).
– When components are replaced it is necessary
to ensure the traceability of the components
replaced.
– If screws with thread locker are loosened,
these must be replaced with new ones.
– If screws with circlip rings are loosened, these
must be replaced with new ones.
– Parts that become damaged during removal
must be replaced with new ones.
4.1 Spare Parts
WARNING!
Original spare parts must be used in all repairs.
Otherwise the warranty and product declaration
of conformity shall be rendered invalid.
All spare parts must be obtained from the Invacare
customer service department. An electronic spare parts
catalogue can be found at www.invacare.eu.com.
4.2 Replacing Side Flaps
Difficulty:
●○○
Tool: —
Pos.
Designation
Article
number
Quantity
A
Side flap
1563646
2
A
B
Fig. 4-1
Removal
1.
Press in the clips of side flap
A
and slowly draw out
the side flap.
Fitting
1.
Slide the clips of side flap
A
evenly into the slots of
the seat plate
B
until they engage.
4.3 Replacing Hand Control
Difficulty:
●●●
Tool:
Pos.
Designation
Article
number
Quantity
A
Hand control
incl. sealing
1590349
1
Removal
1.
Remove battery.
2.
Remove backrest from seat and put backrest on
worktop.
3.
Remove screws of electronic cover.
4.
Carefully lift up electronic cover from electronic box.
5.
Remove sealing.
6.
Disconnect and remove old hand control from circuit
board.
Fitting
A
B
C
D
Fig. 4-2
1.
Put black grommet
A
of hand control into intended
gap
B
of electronic box.
2.
Insert plug
C
of new hand control into connector
D
on circuit board.
3.
Insert new sealing.
1650139-A
7
Содержание Aquatec KOGIA
Страница 15: ...Notes ...