background image

user to set the range they will stay in during measurement.(see 15E below for informa on on seLng 
the range.)   

15E. Gain Range: 

 When Auto Gain Range is set to off in sec on D above, Gain Range becomes 

accessible.  The user can select the range based the max current to be measured. Ie; 1 mA, 3uA or 300
nA.  Each relay allows the ILT2500 to measure approximately 3-4 decades with the max being the 
satura on point and noise become problema c as the reading approaches .0005 from the max.

To determine the range of current that applies to a specific light level reading, the user can select 
current in the display units (see sec on 14A for instruc ons)

 

or mul ply a previous reading by the 

calibra on factor used to obtain the reading. Ie. 1mW/cm2 (1e-3 mW/cm2) with a calibra on factor of
1e-3 (A)(cm2)(W-1) =.000001 amps,  1e-3 * 1e-3.

15F. Min-Max (and FW Peak)

:  The min-max seLngs allows the user to change the func onality of the

min-max bu3on. 

When in Min/Max mode, the reading will only change for Max when a reading that is higher than the 
displayed value is read, and for Min when a reading that is lower than the displayed value is read.  
When FW peak is displayed the min-max bu3on will change to a FW Peak bu3on.  FW Peak uses the 
fastest scanning speeds to see rapid changes in light intensity and uses less averaging. For Very Fast 
moving light sources such as scanning lasers or pulses of light, use FW peak. For be3er averaging we 
recommend using min-max when speed is not a factor.

15G. 5V Bias On or Off:

  The defaults seLng is off as there are numerous sensors that can be 

damaged or read incorrectly if bias is applied.  When turned on, the meter will supply reverse bias to 
any sensor plugged into the device. (See sec on 5 for pin out informa on pertaining to bias)
Silicon sensors such as the SED033, SED100, SPD025Y, and the XRL series can be biased when 
measuring flashing or rapidly changing values to increase response  me.  SED240, SED220 and SED185
Vacuum phototubes require the bias to be turned on to obtain the needed power.

WARNING: Sensors can provide inaccurate readings or be damaged if inadvertently biased. The ILT 
SED005 (GaAsp), and SED623, SED624 and SED625 (thermopiles) should not be supplied with 5V bias.

Recording to File Se$ngs

15H. Recording Interval:  

Ranges from 1 second to 1 hour, or sample for

faster speeds.
Sample: when selected, the record interval and sample  me are synched. 
Ie if sample  me is set .1, the meter will record a value every .1 seconds.

Trend Chart Se$ngs

15I. Chart Type: 

Select a bar graph or line graph

.

15J. Update Interval: 

Select the  me between updates to the display

from .1 sec to 1 hour.

Error and Warning Level SeLngs15K. Light Stack SeLng:  SeLngs for light
stack will determine when the screen turns Red, Yellow, and Green.
Trigger Type:  There are three available triggers to ini ate a change in the

ILT2500 INSTRUMENTATION MANUAL  Preliminary 8/18     12 OF 22

Содержание ILT2500

Страница 1: ...ILT2500 Opera on Manual ILT2500 manual www intl lighttech com 1 of 22...

Страница 2: ...t Measurement and Limita ons 8 9 Sample Time and Limita ons 9 10 5V Bias 9 11 Home Screen 9 12 Display Style Selec on 10 13 Display Content 10 14 Feature Bu ons 11 15 Se5ngs 11 16 Files 16 17 Light Ca...

Страница 3: ...480x272 TFT Color LCD Touch enabled Backlit C Opera ng Temperature 0 40 Deg C D Storage Temperature 10 45 Deg C E Memory 4GB microSD F USB Current Draw Typical 500mA Max 900mA USB 3 0 G Bias Programm...

Страница 4: ...in sensors mfg 2018 or later 2 Turn on the meter by pushing the On O bu3on down into the I posi on 3 A er a short start up menu the Home Screen will appear 4 Check the power monitor window in the upp...

Страница 5: ...mbient zero is used to subtract small background light levels that are not coming from the test source With the test source o or blocked and the sensor exposed to the environment ambient light press z...

Страница 6: ...al temperature can rapidly increase and cause erroneous readings and damage Standard opera ng temperature is rated at 0 400 C Do not apply any force to the display Do not apply pressure to extended st...

Страница 7: ...Analog In 1 Pin6 Det Cathode Pin14 External Analog In 2 Pin7 Det Anode Pin15 N C Pin8 5Volts Warning please consult with an ILT technician before connec ng a current genera ng device that could excee...

Страница 8: ...hen this condi on occurs the display will provide a SATURATION warning To prevent satura on neutral density lters or apertures can be added to the sensor or a greater distance between the sensor and t...

Страница 9: ...material ILT silicon sensors include SED033 SED100 SCD110 XSD140B XSD340B XSD140A and XSD340A 11 Home Screen When the meter is turned on it will take a few moments for the so8ware to ini alize Upon c...

Страница 10: ...e as fast as 1 seconds or as slow as once per hour Note See see 15I 15J for seLng bar or line graph 14 Feature Bu ons 14A Display Units Toggles between the available fields Current output is in amps O...

Страница 11: ...alibra on factors 15B1 Meter Factor Selects calibra on factors from the pre programmed available list ranging from 1 20 Factor selec on allows readings in calibrated units such as W W cm2 Lux etc 15B2...

Страница 12: ...es such as scanning lasers or pulses of light use FW peak For be3er averaging we recommend using min max when speed is not a factor 15G 5V Bias On or O The defaults seLng is o as there are numerous se...

Страница 13: ...3 For High and Range the error level must be greater than the warning level Ie target is 1 mW cm2 1e 3 the warning must be larger than 1e 3 ie 3e 3 or 3 mW and the error must be larger than the warni...

Страница 14: ...ed arrow on cal cert on page 14 15Q 5 Resul ng Light Level Allows the user to enter a know value for the light level under test and create a calibra on factor to force the meter to read the correct va...

Страница 15: ...adings When the ILT2500 is connected to the computer using the USB micro cable the computer will register the ILT2500 as a device with removable storage allowing access into the Files folder To access...

Страница 16: ...the loca on to apply With the box showing the loca on touch the Del delete bu3on The Delete bu3ons border will turn blue to indicate comple on If you want to back out of Delete scroll to the top and...

Страница 17: ...in SeLngs The white display box shows the ash count in the upper le Note Counts was designed for rapid rises in light intensity and increases when the signal rises 4x between 100uS samples the working...

Страница 18: ...generated light level measured exceeds the measurement capabili es of the system Increase the distance add a3enua on lters or add an aperture to the sensor to allow measurement of stronger signals Low...

Страница 19: ...00 decreases the number of digits of strobes to count v is a programmable value of how many ashes to be measured Level selec on High Light level 3 uA to 1 mA Frequency 3 1 4 Hz Low Light 0 to 3 uA Fre...

Страница 20: ...hould be plugged in during ini aliza on Insert the USB s ck into the USB A port on the side of the ILT2500 Click on Settings and Scroll to the bottom of setting window Click on the rmware update box S...

Страница 21: ...pressure to the meter when the stand is in use 28A Opera ng Temperature 0 40 C ILT meters and sensors have a recommended opera ng range of 0 to 40 C to assure the accuracy of the calibrated measuremen...

Страница 22: ...parts or repair by other than Interna onal Light Technologies The equipment will be repaired or replaced at our op on without charge to the owner for parts or labor incurred in such repair This warra...

Отзывы: