5.2 Hardware installation
The rotary chair must be secured to the floor with 4 screws
through the bottom rim of the base. Please use 14 mm dowels
and 10 mm screws.
When orienting the chair in the examination room please observe that the default zero position of the
chair is indicated by the exit point of the cables from the chair.
Never connect the instrument to the power outlet before the side
panels and the top cover have been correctly refitted and
secured.
Consult with the product specification plate on the unit for c
ompliance between the unit‟s own data and
those of the local power supply system (mains voltage and mains frequency) before connecting the
unit.
Connect only if all data agrees!
Inspect the mains connection cables for visible damage prior to establishing the connection. Damaged
cables or plugs must immediately be replaced by Interacoustics service technicians, an authorized
person or specialist.
Connection Panel
The backrest features a connection panel with purpose specific connections:
2 x 6-pM FireWire ports
1 x USB 2.0 port
1 x DC out (not active)