Rotary Chair Nydiag 200
Page
6
Thus, the units should be effectively protected against moisture. Ventilations lots should be kept
unobstructed for maintaining circulation of air.
The instruments should generally be stored at temperatures between 0° to +50°C/+32° to 122°F.
3.4 Hygiene
Always be sure to pull the plug before cleaning or disinfecting mains operated instruments.
The instrument must be cleaned with a non-fuzzy, slightly
moistened piece of cloth (please do not soak it!) Only use a mild
alcoholic fluid as a disinfectant.
When cleaning the instrument, any moisture, like e.g. condensation water, must necessarily be
prevented from getting inside!
Chemicals required for operation or care of the unit must always be stored, prepared and made
available in specially labelled containers to prevent any mistakes.
3.5 Maintenance
Before switching on the instrument make sure that the mains cable, mains plug, socket and input at
the instrument are in a perfect state. If the drive of the rotary chair becomes noisy the driving belts
should be renewed. We recommend renewing the driving belts after five years.
The operator is responsible for the perfect state of the instrument. It is recommended by
Interacoustics to have safety checks carried out at one-year intervals.
Immediate maintenance is required if:
the instrument was subjected to extreme mechanical stress (impact, defective cable due to
inadmissible traction);
fluid gets into the instrument;
cables and/or connectors are defective;
rubber-joints have developed cracks.
For further details please contact your Interacoustics representative.
S
YMBOLS ON
I
NTERACOUSTICS COMPONENTS
Ground conductor connection
Unit of type B