Rotary Chair Nydiag 200
Page
5
Interacoustics equipment is in compliance with Protection Category I
Connection of equipment of type BF inside or outside medically utilized locations
Due to the increased safety requirements, medically applied instruments must not be connected via an
extension flex or a multiple socket outlet.
The instrument must exclusively be connected via a wall
socket with a ground contact!
Never connect the instrument to a power outlet before all
panels and fittings have been securely mounted
Coupling of equipment to other systems
External equipment intended for connection to signal input, signal output or other connector, shall
comply with relevant IEC standard (e.g. IEC 60950 for IT equipment and the IEC 60601 series for
medical electrical equipment). In addition, all such combinations - systems - shall comply with the
standard 60601-1-1, Safety requirements for medical electrical systems. Equipment not complying with
IEC 60601 shall be kept outside patient environment, as defined in standard (at least 1.5 m from the
patient).
Any person who connects external equipment to signal input, signal output or other connectors has
formed a system and is therefore responsible for the system to comply with the requirements of IEC
60601-1-1. If in doubt, contact your service technician or local representative for help
Should any further information be required, please do not hesitate to contact your Interacoustics
representative.
3.3 Ambient conditions
Interacoustics equipment has not been conceived for operation in medically utilized rooms with
potential explosion hazards.
The equipment should not be installed in the vicinity of e.g. X-ray equipment, motors or transformers
with high installed power rating, since electric or magnetic interferences may falsify the results of a
measurement or even make measurements impossible.
Therefore, the instrument should not be installed in the vicinity of power lines.
If not otherwise stated in the shipping documents, Interacoustics equipment in the normal encased
version has been conceived for operation under normal climatic ambient conditions.
Temperature:
+ 15°C/+59°F
…
+ 35° C/+95°F
Relative humidity:
30
…
90 %
Barometric pressure:
500
…
1060HPA