background image

 

D-0128226-A – 2020/12 
Air Fx Caloric Irrigator – Instructions for Use - US 

 

Page 

9

 

 

 

Figure 3 

Place Speculum on Otoscope Head 

 

Figure 4

 Turn Speculum Clockwise to Lock, Pull Gently on 

Speculum to Verify It’s Secure 

 

Specula are latex free and silicone free. Do not put too much torque on the speculum or the plastic nub 

on the inside of the speculum will break, preventing the speculum from securing to the handle. Once this 

tip is broken, the speculum cannot be secured again on the handle. The speculum can be removed from 

the handle by first turning the speculum counterclockwise to unlock. 

The speculum must be tight to 

prevent air leaks that will reduce the caloric stimulation response.  

 

 

3.5  Select the irrigation temperature  

If the Air Fx is used as a standalone device, press the Cool or Warm button on the front panel to select 

the irrigation temperature. If the Air Fx is connected to a VNG/ENG system with compatible software, then 

the software will select the irrigation temperature based on the test or system settings.  

 

NOTICE: If the irrigator system is used along with VNG/ENG software, refer to the respective software 

user manual for computer specification and supported operating system related information.  

 

The irrigator front panel LCD display will show the irrigation method (Cool or Warm) followed by the 

desired irrigation temperature (24°C) and the current temperature (e. g. 23. 4°C).  

Cool:

 

24°C

 

23. 4°C

 

 

 

  

  

 

Irrigation will be available after the desired temperature has been reached and stabilized. Once this 

occurs, the Air Fx will emit an indicator tone and will display “

Ready

” below the irrigation information.  

Cool:

 

24°C

 

24. 2°C

 

Ready 

 

  

  

 

 

3.6  Perform the irrigation 

 

An audiologist or physician should inspect the patient’s ear with an otoscope 

prior to testing, looking for infection, open wounds, wax impaction, or a 

perforated eardrum. If any of these are observed, DO NOT proceed with 

irrigation using the Air Fx irrigator. 

 

Prior to irrigation, the patient should be supine with the head elevated 30 degrees to place the lateral 

semi-circular canals in the vertical plane. The operator should sit next to the patient alongside the ear 

being irrigated. Hold the irrigator handle in one hand and instruct the patient what will occur and what the 

patient will feel. Insert the otoscope with the 2. 75mm diameter speculum into the patient’s ear while 

peering through the otoscope magnifying lens. Adjust the direction of the speculum looking for the 

tympanic membrane (TM). Direct the air stream continuously at the TM during the whole irrigation. Press 

the white push button on the irrigator handle to start the countdown timer. If connected to the computer 

with compatible VNG/ENG software, the test will begin recording as well. During irrigation the LCD 

display will display “Irrigating” and the amount of time that has elapsed. To restart the test hold down on 

the white button for 3 seconds and the timer will reset and abort the VNG/ENG tracing.  

 

 

Содержание Air Fx

Страница 1: ...Instructions for Use US Air Fx Caloric Irrigator Science made smarter 8525533 D 0128226 A 2020 12...

Страница 2: ...thout notice No part of this document may be reproduced or transmitted in any form or by any means without a prior written permission of Interacoustics A S FireWire is a registered trademark of Apple...

Страница 3: ...re 9 3 6 Perform the irrigation 9 3 7 Safety 10 3 8 Turn off the irrigator 10 3 9 Draining the water reservoir 10 4 MAINTENANCE 12 4 1 General cleaning procedure 12 4 1 1 General Precautions 12 4 1 2...

Страница 4: ...the caloric test with the correct irrigation temperature All personnel who operate the Air Fx should familiarize themselves with the contents of this manual prior to using the Air Fx with a patient A...

Страница 5: ...C 10 Amperes Speculum 2 75 mm Speculum 4 25 mm 60cc syringe and tube USB Cable 3m 9 8ft User s Manual Drain Fill Kit Hose Management Tape 1 4 Warnings and precautions Throughout this manual the follow...

Страница 6: ...he invoice serial number and a detailed report of the issue For any on site service related information please contact your local distributor If the system components are to be returned for service pl...

Страница 7: ...sed at close distances to the equipment Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment Performance Conditions...

Страница 8: ...ons for Use Observe precautions for handling electrostatic sensitive devices It is against the law to dispose of electronic devices in the trash The crossed out wheelie bin symbol shows that the compo...

Страница 9: ...ol Irrigation to be performed returns to standby state C Adjust temperature set point up 1 C D Adjust temperature set point down 1 C E Hose connection F Handle support G LCD output screen Figure 2 Rea...

Страница 10: ...e tube to the Fill Drain Water Outlet Fully push the tube into the Fill Drain Water push in connector about 1 2cm of tubing Gently pull on the tubing to confirm the tube is secure 5 Switch the Fill Dr...

Страница 11: ...urn on the irrigator Turn on the power switch on the back panel The Air Fx will initialize show the firmware version and then go into the standby state Select Cool Warm Air Fx v1 4 3 4 Attaching the s...

Страница 12: ...mperature e g 23 4 C Cool 24 C 23 4 C Irrigation will be available after the desired temperature has been reached and stabilized Once this occurs the Air Fx will emit an indicator tone and will displa...

Страница 13: ...n the irrigator and isolate the problem ambient temperature too low bubbles in the water line distilled or demineralized water level too low etc Contact Interacoustics Service Department or the local...

Страница 14: ...ube Squirt any distilled or demineralized water into a waste container Reconnect the syringe to the tube Repeat this process until all of the distilled or demineralized water is removed from the inter...

Страница 15: ...ion control The WHO s recommended level of decontamination for non critical devices is cleaning Disinfection and sterilization are not recommended However in case of an epidemic outbreak disinfection...

Страница 16: ...r Then fill the irrigator with distilled or demineralized water and run couple of cool irrigation cycles Drain the irrigator to ensure that the disinfectant is washed out Then refill the reservoir wit...

Страница 17: ...ed Interacoustics service representative altered in any way so as in Interacoustics judgement to affect its stability or reliability subject to misuse or negligence or accident or which has had the se...

Страница 18: ...ctric and electronic waste as unsorted municipal waste Electric and electronic waste may contain hazardous substances and therefore has to be collected separately Such products will be marked with the...

Страница 19: ...D 0128226 A 2020 12 Air Fx Caloric Irrigator Instructions for Use US Page 16...

Страница 20: ...faster 3 m 9 8 ft cable Case dimensions 35W x 32D x 22H cm 13 8 x 12 6 x 8 7 in Air hose length 3 m 9 8 ft Weight 8 kg 18 lb Voltage 110 130 VAC 50 60 Hz 220 240 VAC 50 60 Hz Fuses 2 each 110VAC unit...

Страница 21: ...ons or decreased electromagnetic immunity of this equipment and result in improper operation The list of accessories and cables can be found in this section Portable RF communications equipment includ...

Страница 22: ...omagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Air Fx can help prevent electromagnetic interferences by maintaining a minimum distance between p...

Страница 23: ...nvironment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or residential environment Vol...

Страница 24: ...meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the...

Страница 25: ...s or fax No to whom Interacoustics may confirm reception of the returned goods The above mentioned item is reported to be dangerous to patient or user 1 In order to ensure instant and effective treatm...

Отзывы: