4.3 TROUBLE SHOOTING

!!  CAUTION:

The electronics are not protected against condensed

liquid water inside the enclosure.  If conduit is used, be sure conduit is dry
or sealed at the instrument to prevent conduit condensation from entering
the enclosure.

TABLE I.  TROUBLE SHOOTING GUIDE

OBSERVATION

PROBABLE CAUSE

REMEDY

Flow trip level continually
drifting downward with constant
flow.

1. Coating forming on

wetted surface of
transducer.

1. Clean transducer

periodically.

2. Adjust trip potentiometer

clockwise until layer
build-up stabilizes.

After switch has been operating
properly: relay trips with flow
above trip level and cannot be
adjusted using SECTION 2.4
instructions.  Relay does not trip
when flow falls below trip level
and cannot be adjusted using
SECTION 2.4 instructions.

1. Bad electronic

component.

1. Contact factory.

Relay cannot be made to trip by
adjusting "Trip" potentiometer.

1. Initial flow rate estimate

was too low or too high.

2. Flow media change.

1. Check flow conditions.
2. Contact factory.

LED is not lit and relay stays in
alarm condition (N.C. contact is
made).

1. No power to electronics.
2. Blown fuse.

1. Check incoming power.
2. Replace fuse with ¼

Amp fast acting fuse
(fuse is in PWB socket).

3. Contact factory. 

Содержание Rheotherm 100CS

Страница 1: ...have been thus defective this warranty being expressly in lieu of all other warranties express or implied and all other obligation or liabilities on Intek s part The purchaser will assume all responsi...

Страница 2: ...START 7 3 2 SWITCH SETTINGS 7 3 3 GENERAL INFORMATION 7 3 4 USE AS A LEVEL SWITCH 7 SECTION 4 MAINTENANCE 8 4 1 GENERAL MAINTENANCE 8 4 2 SPARE PARTS 8 4 3 TROUBLE SHOOTING 9 TABLE I TROUBLE SHOOTING...

Страница 3: ...ndicating high and low flow conditions Key features of Rheotherm flow switches are Non intrusive flow monitoring For pipe sizes from 030 to inch flow sensing is done from outside the flow tube No movi...

Страница 4: ...ive transducer Example of Example of TU probe with probe with NPT fitting flange fitting 1 3 PRECAUTIONS 1 Use proper input power Check the label on the unit for the input power requirements 2 Use rea...

Страница 5: ...nstructionscoverinstallation operationandmaintenance ofRheothermflowswitches instandard configurations Any special information pertaining to your unit is covered under CUSTOM INFORMATION SECTION 6 Tim...

Страница 6: ...alling CAUTION The electronics box should never be rotated for any reason Straight run for a flow switch is not a requirement particularly for flow no flow indications but for best repeatability some...

Страница 7: ...el shall be properly grounded with an anti static device to prevent damage to components within the electrical enclosure The input power requirement is listed on the tag on the electronics enclosure m...

Страница 8: ...0 seconds and depends on where the trip adjust is set 2 4 ADJUSTING THE TRIP POINT Adjust the flow switch trip point as follows see Figure 2 for trip potentiometer location All adjustments must be mad...

Страница 9: ...O contact is open and LED color becomes red High Flow Switch The relay operation is exactly as described above When used as a high flow switch the alarm condition is considered to be when the relay i...

Страница 10: ...nts gas bubbles and non solubles can result in measurement below actual flow rates Periodic checks and cleaning should be performed to insure a clean pipe or probe surface It should be part of normal...

Страница 11: ...iometer clockwise until layer build up stabilizes After switch has been operating properly relay trips with flow above trip level and cannot be adjusted using SECTION 2 4 instructions Relay does not t...

Страница 12: ...ewed SECTION 4 3 TROUBLE SHOOTING and have questions please call our experienced engineers for assistance In many cases we can solve a problem over the phone Please provide as complete a description a...

Страница 13: ...CATION Model no Serial no Customer identification 6 2 CONFIGURATION The configuration of this unit as originally shipped from the factory Input Power 24 Vdc Line Connection 6 3 SPECIAL INSTRUCTIONS Re...

Страница 14: ...terface connector Enclosure lid 24 14 Ga Flow switch electronics Check enclosure label for proper input power For VDC input connect positive to H or terminal and negative to N or Figure 1 Model 100CS...

Отзывы: