SECTION 4 

!

 MAINTENANCE

4.1 GENERAL MAINTENANCE

Certain precautions should be taken to insure proper performance of all models of flow instruments.
Since the measurement technique involves a signal resulting from heat transfer to the flowing medium,
care should be exercised to prevent build-up of varying layers on the walls of the transducer.  Layers
such as bacterial growth, dried paints, gas bubbles and non-solubles can result in measurement below
actual flow rates.  Periodic checks and cleaning should be performed to insure a clean pipe or probe
surface. 

It should be part of normal maintenance procedure to check the system for proper functioning.
Experience and other observable conditions should be utilized to determine the frequency of inspection.
Long term drift in the unit calibrations is not expected, but if a single point recalibration is required,
refer to SECTION 4.2.  To test the flow switch action (if that option was ordered), the flow rate should
be reduced below (for low flow switch) or raised above (for high flow switch) the switching level.  Then
check and insure relay action and continuity of the shut down or warning circuits which it operates. 

The joints of all intrusive probes tips should be inspected for wear and corrosion.  

4.2 SINGLE POINT FIELD CALIBRATION 

When field calibration is required, for any reason, the simplest adjustment is a single point linear offset,
and this should usually be tried first.  All calibrations should be done at normal operating temperature.
When a single point calibration (or recalibration) is required, it is performed as follows:

Establish a known flow rate in the upper third of the calibrated range that is shown on the output
curve (Figure 2).  Adjust calibration potentiometer P2 (cal), Figure1, for a correct output reading
or signal.

If P2 runs out of adjustment before the meter is reading correctly, P1 (see Board Layout P.17) can be
used to further offset the signal.  Turning P1 clockwise decreases the signal.  (Note:  P1 has no effect
on some meters.)

Sometimes you may find the single point calibration is not sufficient to align the output over a wide flow
rate range.  If so, a two-point offset/gain calibration can be performed as described below. 

A.

Establish a known flow rate at the low end (10-15% of full scale) of the factory calibrated
range (Figure 2, page 18).

1

  Adjust P2 potentiometer, Figure 1, for an accurate output signal.

B.

Establish a flow rate near 90-95% of full scale.

1

  Adjust "Flow Span" potentiometer, P3

(Figure 1) for an accurate output signal.

C.

Repeat steps A and B until desired accuracy is achieved.  Record flow vs. output relationships
for various flows and compare with output curve.  Construct a new calibration curve if needed.

1

The flow meter output must have stabilized before any adjustment is made.  A strip chart or data logger

monitoring of the output can be v ery helpful in determining when the flow has entered steady-state conditions,
especially at low or noisy flows.

Содержание Rheotherm 100

Страница 1: ...isclose to its satisfaction to have been thus defective this warranty being expressly inlieu of all other warranties express or implied and all other obligation or liabilities on Intek s part The purc...

Страница 2: ...al Check 11 SECTION 4 MAINTENANCE 12 4 1 GENERAL MAINTENANCE 12 4 3 SPARE PARTS 13 4 4 TROUBLE SHOOTING 13 TABLE I TROUBLE SHOOTING GUIDE 13 SECTION 5 CUSTOMER SERVICE 14 5 1 QUESTION ON EXISTING HARD...

Страница 3: ...ures of Rheotherm instruments are Nonintrusive flow measurement For pipe sizes from 0 030 to 2 inches flow sensing can be done from outside the flow tube No moving parts There are no rotating translat...

Страница 4: ...non linear output is approximately proportional to the logarithm of flow Intek Inc is licensed to use this patented and trademarked flow measurement technique Nonintrusive transducer Example of Exampl...

Страница 5: ...clean and free of permanent layer build up 7 Do not exceed pressure limits of the tube or fittings 8 Maintain a thermally stable environment short term for the transducer and adjacent line See SECTION...

Страница 6: ...and if supplied the gasket is in place Seal conduit lines at the connector box if conduit lines can become wet 1 TU nonintrusive TU 1 16 and TUc transducers unless they have optional O D ends particul...

Страница 7: ...uniform fluid temperature A box around the sensor and inlet tubing is highly recommend for operating temperatures higher than room ambient Allow enough inlet tubing inside the box to allow the fluid...

Страница 8: ...ronics housings are available These include NEMA 4 laboratory bench type explosion proof and panel mount enclosures as well as special models to meet customer requirements These come in different size...

Страница 9: ...rly after installation is completed check these connections again In general the laboratory unit will have a plug in connector for the transducer cable 2 Power The input power requirement is listed on...

Страница 10: ...r a particular fluid either at the factory or in the field If the fluid changes properties the calibration changes Therefore once calibrated do not allow fluid properties such as density and viscosity...

Страница 11: ...2 Transducer isolation check Figure B p 11 3 Electronic circuit board check Figure C p 11 1 The transducer continuity check is performed as follows see Figure A p 11 A Disconnect the transducer cable...

Страница 12: ...the TEMP and the GND test point jumpers Select R3 for proper TEMP value Select R2 coarse and adjust R1 fine for T value G Compare the measured flow rate and the indicated flow rate to original calibra...

Страница 13: ...11 MODEL100 REV D wpd...

Страница 14: ...is a single point linear offset and this should usually be tried first All calibrations should be done at normal operating temperature When a single point calibration or recalibration is required it i...

Страница 15: ...izes Output saturates high or low will not respond to flow change 1 Loose transducer cable connections 2 Flow rate not yet within range of meter 3 Meter Calib switch in wrong position 4 Transducer cab...

Страница 16: ...the phone Please provide as complete a description as possible of the problems encountered 5 3 FACTORY AND FIELD SERVICE If you request field service Intek has experienced engineers available to meet...

Страница 17: ...0 Vac 50 60 Hz Q 24 Vdc Output 4 20 mA Q 0 5 Vdc Q 0 10 Vdc Pressure Rating Temperature Rating Calibrated Range Wetted Material 316 stainless steel Q other Enclosure Type NEMA 4 Q NEMA 4X Q NEMA 7 CL...

Страница 18: ...TA FOR FUNCTIONAL TEST Unit Serial Number I II III IV V VI TEMP T TP4 Linear Output Instrument Display Flow Rate Note An offset of data in Column III and IV with respect to Column II may appear if the...

Страница 19: ...17 Figure 1 Rheotherm 100 Rev B Signal Conditioner PCB Layout...

Страница 20: ...18 Figure 2 Flow Rate vs Output Curve...

Отзывы: