- 10 -
MODEL100 REV D.wpd
E.
Restore power.
F.
Connect a DC voltmeter across the TEMP (
%
) and the GND (
!
) test point jumpers.
Select R3* for proper TEMP value. Select R2* (coarse) and adjust R1* (fine) for
)
T
value.
G.
Compare the measured flow rate and the indicated flow rate to original calibration data
of Table II. Recheck TEMP and
)
T.
*resistors of dummy transducer