INSTRUCTIONS FOR USE
8
RESETTING THE AXOR™ II AFTER RELEASE IN BENDING/FLEXION
When the Axor™ II has released in bending/flexion direction it can be reset:
1.
Sit on a chair or on the floor with the leg straight out and the heel of the prosthetic foot on a solid support. If
needed, remove the prosthesis from the Abutment before resetting the Axor™ II.
2.
Gently close the Axor™ II until the green release plunger makes contact with the edge of the housing.
3.
Press down on the Axor™ II according to Figure 7 until it clicks back into the correct position. A moderate to high
force may be required.
CAUTION!
Avoid resetting the Axor™ II if there is any pain during the reset procedure. In case of pain, please contact
your physician immediately. Please also contact your prosthetist for reset of the Axor™ II.
CAUTION!
Try to keep fingers away from the bend release mechanism during reset to avoid pinching the fingers.
RESETTING THE AXOR™ II AFTER RE-
LEASE IN ROTATION
When the Axor™ II has released in rotation it is reset by
turning the Axor™ II until the prosthetic leg is back in the
normal position for walking (see Figure 8). A click is felt
when the release mechanism is correctly reset.
CAUTION!
A similar click is also felt when the leg is
rotated 180 degrees, i.e. when the foot points straight
backwards. If this happens, apply a force to turn the foot
into the correct position.
Figure 7. Resetting the Axor after release in bending/flexion.
Figure 8. Resetting the Axor™ II after release in rotation.
Содержание OPRA AXOR II
Страница 1: ...OPRA AXOR II Instructions For Use ...