FIREBOY eco - Operating instructions V07
Operation
18
INTEGRA Biosciences AG
Engl
ish
.
Error message Technical cause
Possible cause and correction
Red LED is
flashing
(1x per sec),
green LED is on.
The flame monitor
does not detect
a flame.
The valve is closed.
No gas supply is connected.
Connect the gas supply
Open the gas connection
Flame was blown out several times.
Correctly set the gas/air mixture
Prevent draft
Electrode of the flame monitor is
broken or dirty.
Visually inspect electrode
Clean electrode with alcohol
Red LED is on,
green LED is on.
Prior to ignition
process, the contact
pin on the burner head
is not detected.
Burner head not added or not added
correctly.
Add burner head
Contact pin on burner head
is contaminated.
Clean the contact pin
Red LED
is flashing
quickly
(5x per sec),
green LED is on.
Burner chamber
temperature
measured exceeds
the critical value.
Continuous operation overheated
the device.
Wait until device has cooled off
Ambient temperature exceeds
the permitted value.
Reflection of the flame on the device
due to unauthorized use.
Red LED
is flashing slowly
(every 3 sec),
green LED is on.
Failure of
flame monitor.
Electrode of the flame monitor
is kinked or dirty.
Visually inspect, if necessary
clean
Red LED is on,
green LED is off.
Technical error.
Call service technician