FIREBOY eco - Operating instructions V07
Installation
12
INTEGRA Biosciences AG
Engl
ish
When replacing the nozzle, proceed as follows:
3.2.2 Installation and removal of the adapters for the gas connection
The adapters for the central gas supply, the gas cylinder, or the gas cartridge
are inserted in the gas adapter opening of the FIREBOY eco without the gas
cartridge (see “Rear view” on page 10).
Insert the appropriate adapter in the gas adapter opening of FIREBOY eco,
making sure it audibly clicks into position.
The adapter is now installed.
W
ARNING
When replacing the adapters, always switch off the device, disconnect the
mains plug, and close off the gas supply.
Unscrew the brass screw (1) on the lower
surface of the device using a small coin
or screwdriver.
Turn over the device, allow the nozzle to drop
out, and insert the desired new nozzle (2).
Retighten the brass screw to the stop.
The nozzle is now inserted.
1
2
In order to remove the adapter, use
a pointed tool to press the adapter
mechanism inward on the lower surface
of the device and remove the adapter
by pulling toward the rear.
Insert the requested new adapter until
it audibly clicks into position.
The adapter is now replaced.