85
7
If
BG sound
is set to
On
, the PDM beeps twice. The PDM
automatically marks the result as a control solution test (Fig-
ure 7-8 on the previous page).
7. Compare the control solution test result to the range printed
on the Test strip vial label. If the result does not fall within the
range, repeat the test. If the result still does not fall within
the range, call Insulet’s Customer Care (24 hours/7 days):
800-591-3455. When calling from outside the United States:
781-457-5098.
8. Remove the test strip. If the strip port light was turned on, it
now turns off.
If you remove the test strip before this step, you will still see
the available soft keys.
n
Performing a Blood Glucose Reading
Figure 7-9 shows the parts of the lancing device.
If the words
Control solution
do not show on
the screen, you MUST tag this result as a control
solution test.
If you do not mark this result as a control solution
result it will be included in your BG averages and
your averages may change, too. Repeat the test
with a new test strip. If the problem persists, con-
tact Insulet's Customer Care (24 hours/7days):
800-591-3455. When calling from outside the
United States: 781-457-5098.
Be sure to discard used test strips. (You can only use
test strips once.) Used test strips are considered bio-
hazardous material.
Release Button
Depth Wheel
Cocking Control
Figure 7-9
Lancing Device Cap
Depth Indicator
Ejection Control
Lancing device depicted in Figure 7-9 is representative only. Please follow
the instructions included with your specific lancing device.