The following table explains what the message could mean and what you should do.
Alert error message
Possible causes
Action
BG Meter Error 1.
Meter problem.
For customer support
call 800.591.3455.
•
There is a problem with the
meter.
1. DO NOT use the meter. Contact Customer Care.
BG Meter Error 2.
Meter or strip
problem.
Retest with a new
strip.
•
Error message could be
caused either by a used
test strip or a problem
with the meter.
1. Repeat the test with a new test strip.
2. If the error persists, call Customer Care or repeat the blood test.
BG Meter Error 3.
Meter was not
ready.
Retest with a new
strip.
•
The sample was applied
before the meter was
ready.
1. Be sure you see the Blood Drop and Test Strip symbols on the display
screen before you apply blood or control solution.
2. Conduct a control solution test using a new test strip. If the results of
the control solution test are within the range printed on the side of
the test strip vial, retest using blood and a new test strip.
3. If the control solution test does not work or the error persists, call
Customer Care or perform another blood test.