![Insulet mylife OmniPod Скачать руководство пользователя страница 77](http://html1.mh-extra.com/html/insulet/mylife-omnipod/mylife-omnipod_manual_2070293077.webp)
Using the Personal Diabetes Manager
65
6
a. To edit an existing
target BG
time segment, choose the
segment, then press
Edit
. Use the
Up/Down Controller
buttons to choose a new start time, end time, target BG
value, and correct-above value (
correction threshold
). Press
Next
after each entry.
b. To add a segment, choose
[add new]
, then press
New
.
Use the
Up/Down Controller
buttons to choose a start
time, end time, target BG value, and correct-above value
(correction threshold). Press
Next
after each entry.
c. When you have completed all the time segments you
want, press
Done
, then
Save
.
2.
Min BG for calcs
(smallest blood glucose value allowed for
calculating boluses)—Use the
Up/Down Controller
buttons
to enter a new minimum BG, then press
Next
.
3.
Insulin-to-carb (IC) ratio
—You can edit existing time seg-
ments and add segments, up to a total of 8.
a. To edit an existing
IC ratio
time segment, choose the seg-
ment, then press
Edit
(Figure 6-7 on next page). Use the
Up/Down Controller
buttons to choose a new start time,
end time, and IC ratio. Press
Next
after each entry.
b. To add a segment, choose
[add new]
, then press
New
.
Use the
Up/Down Controller
buttons to choose a start
time, end time, and IC ratio. Press
Next
after each entry.
c. When you have completed all time segments, press
Done
, then
Save
.
4.
Correction factor
—You can edit existing time segments and
add segments, up to a total of 8.
a. To edit an existing
correction factor
time segment, choose
the segment, then press
Edit
. Use the
Up/Down Control-
ler
buttons to choose a new start time, end time, and cor-
rection factor. Press
Next
after each entry.
b. To add a segment, choose
[add new]
, then press
New
.
Use the
Up/Down Controller
buttons to choose a start
time, end time, and correction factor. Press
Next
after
each entry.
c. When you have completed all time segments, press
Done
, then
Save
.
Figure 6-7
Содержание mylife OmniPod
Страница 147: ...Alerts and Alarms 135 10...