Instrutech CVM211 Скачать руководство пользователя страница 7

 

 

Instruction Manual 

InstruTech CVM-211 "

Stinger

" Convection Vacuum Gauge Module

 

 

 

 

 

InstruTech, Inc 

 page 7  

 

 

 

Setup and Operation

 

 

User Interface Basics: 
 
The user interface is designed for easy operation and a natural 
progression of setup parameters.  A complete user interface map  
is shown in appendix A.  This section gives a brief explanation of 
operation for added clarity. 
 
The CVM-211 Stinger module has 4 settings that can be adjusted by the 
user with the 3 position switch on the side of the module housing.  
Pressing the switch straight in is referred to as pressing the <select> key.  
Pressing the switch upward is referred to as pressing the <up> key.  
Pressing the switch downward is referred to as pressing the <down> key.  
During setup, and operation, be sure to consider the Torr/mTorr LEDs. 
 
Setting up the CVM-211 Stinger Module 
 
1.    With the CVM-211 in the normal pressure display mode, press and hold <select> for 3 seconds.   
2.    The readout displays the value of the '

setpoint turn-on’

 pressure.  The relay energizes when pressure drops  

 

 

below

 this value.  [Factory default = 100 mTorr.] 

3a.  To keep this value and proceed to the next step, press <select>. 
3b.  To change the value, use the <up> <down> keys.  Then press <select> to save and go to the next step. 
4.    The readout displays the value of the 

'setpoint turn-off’

 pressure.  The relay de-energizes when the pressure  

 is 

above

 this value.  [Factory default = 200 mTorr.] 

5a.  To keep this value and proceed to the next step, press <select>.  
5b.  To change the value, use the <up> <down> keys.  Then press <select> to save and go to the next step. 
6.    The readout will display ‘000’ to indicate the unit is in the 

"zero adjust"

 mode.  To properly set "

zero

", the 

 

gauge must be evacuated to a pressure below 10

-4

 Torr.  [Factory default = 000 mTorr.] 

7a.  If the gauge is 

not

 evacuated to a pressure below 10

-4

 Torr, press <select> to proceed to the next step,  

 

without

 saving a new "

zero

" value. 

7b.  If the gauge 

is

 evacuated to a pressure below 10

-4

 Torr, press <down> to save the new user "

zero

"  

 

and proceed to the next step. 

8.   The readout will display the current "

span

" value.  To set "

span

", the gauge must be backfilled with air or  

 

N

2

 to some known pressure above 400 Torr.  (If barometric pressure is known, the gauge may be opened  

 

to atmosphere.)  [Factory default = 760 Torr.] 

9a.  If you do 

not

 have a known pressure in the gauge, press <select> briefly (less than 3 seconds) to exit the 

 

setup menu and return to the normal pressure display 

without

 saving a new "

span

" value.  

9b. If you 

do

 have a known pressure in the gauge, use the <up> <down> keys to change the displayed value to 

 

agree with the known pressure.  Press and hold <select> for 3 seconds until the displayed pressure switches 

 

to the new value.  This will save the new "span" setting, and return to the normal pressure display. 

 
 
Return to Factory Default Settings 
 
You can reset all values to the original factory default settings by holding the <up> key for 5 seconds.  The display 
will read "dEF" until 5 seconds has passed, at which point 

all user settings will be replaced by the original factory 

default values

, and the display will return to the normal pressure display.  If you release the <up> key before 5 

seconds has passed, the display will return to normal pressure display without resetting to factory defaults. 
 

Содержание CVM211

Страница 1: ...11 S ST TI IN NG GE ER R Convection Vacuum Gauge Module User s Manual Revision 1 05 Sept 25 2007 1475 S Fordham St Longmont CO 80503 Phone 303 651 0551 Fax 303 678 1754 E mail info instrutechinc com w...

Страница 2: ...4 Gases other than N2 5 Flammable Explosive Gases 5 Installation Mechanical 5 Position 5 Vibrations 5 Fittings 5 Electrical 6 Connector Pinouts 6 Setup Operation User Interface Basics 7 Setting up th...

Страница 3: ...Torr to 1 0 mTorr 1 digit LED from 0 9 mTorr to 0 1 mTorr materials exposed to gases gold plated tungsten 304 316 stainless steel glass nickel Teflon internal volume 26 cm3 1 589 in 3 internal surface...

Страница 4: ...parts are specified by Instrutech Inc Unauthorized substitutions or non qualified parts may result in fire electric shock or other hazards To reduce the risk of fire or electric shock do not expose t...

Страница 5: ...ure difference error between your process and the gauge and cause a delay in response to pressure changes Mounting the CVM too close to a gas source will also cause errors in the readings Don t mount...

Страница 6: ...practice is to remove power from any cable prior to connecting or disconnecting it The InstruTech CVM 211 will directly replace GP Mini Convectron modules that have a 9 pin D connector and you can us...

Страница 7: ...to save and go to the next step 6 The readout will display 000 to indicate the unit is in the zero adjust mode To properly set zero the gauge must be evacuated to a pressure below 10 4 Torr Factory de...

Страница 8: ...il due to chemical reaction are generally not salvageable Sensors that fail due to coating may possibly be restored by cleaning See Appendix C for a more complete discussion of contamination and clean...

Страница 9: ...hrough Friday at 303 651 0551 Or e mail us at info instrutechinc com If it becomes necessary to return the unit to InstruTech for service or calibration please call or e mail us for a return authoriza...

Страница 10: ...ceed to next step without changing the zero value press select Readout displays current span value Factory default 760 Torr Is system backfilled with air or N2 to some known pressure above 400 Torr Or...

Страница 11: ...of 0 375 to 5 659 VDC for 0 to 1000 Torr of N2 roughly in the shape of an S curve as shown at right GP adopted the same output curve for most of their Mini Convectron modules and controllers with non...

Страница 12: ...r as thermal conductivity goes but even these two gases will exhibit slight differences in readings at higher pressures For example when venting a system to atmosphere using N2 you may see readings ch...

Страница 13: ...essure is 100 Torr the gauge will read 100 Torr If the gas is Ar when the true pressure is 100 Torr the gauge will read only 10 Torr If you are backfilling your vacuum system with Ar when your system...

Страница 14: ...70 1 3150 1 2910 1 2650 1 0020 1 3920 500 mTorr 1 6833 1 3860 1 5890 1 6640 1 6680 1 1940 1 8260 1 8050 1 9140 1 4690 2 0140 1 Torr 2 2168 1 8180 2 1640 2 1950 2 1720 1 5360 2 2570 2 2470 2 6030 1 976...

Страница 15: ...288 5 233 4 845 5 210 5 220 5 616 5 201 5 522 5 0000 5 699 5 389 6 130 5 697 5 524 5 107 5 389 5 418 7 391 5 719 5 877 10 0000 6 000 5 602 8 041 6 013 5 696 5 250 5 471 5 530 8 041 6 332 6 446 20 0000...

Страница 16: ...auge Module InstruTech Inc page 16 Log Linear Analog Output Voltage vs Pressure 0 1 2 3 4 5 6 7 8 9 0 0001 0 001 0 01 0 1 1 10 100 1000 Pressure Torr Output Signal Volts This is a plot of the formulas...

Страница 17: ...87 359 781 1 Torr 1 00 600 940 970 950 400 1 05 1 04 1 55 745 1 60 2 Torr 2 00 1 14 2 22 1 94 1 71 700 1 62 1 66 4 13 1 59 3 33 5 Torr 5 00 2 45 13 5 4 98 3 34 1 28 2 45 2 62 246 5 24 7 53 10 Torr 10...

Страница 18: ...s steel or ceramic to the gases There is no material that is universally chemical resistant your choice of vacuum gauge as well as all other vacuum components should take into consideration the potent...

Страница 19: ...s such as pump oil However depending on your particular contamination you may need to use a more aggressive solvent Cleaning procedure 1 With the gauge port up pour your cleaning solution into the gau...

Страница 20: ...Furthermore the warranty does not apply to products that have been contaminated or when the product or part is damaged during the warranty period due to causes other than ordinary wear and tear to th...

Отзывы: