Instrutech CVM211 Скачать руководство пользователя страница 12

 

 

Instruction Manual 

InstruTech CVM-211 "

Stinger

" Convection Vacuum Gauge Module

 

 

 

 

 

InstruTech, Inc 

 page 12  

 

 

 

 
Effect of Different Gases on Signals and Displayed Pressure

 

A thermal conductivity gauge senses heat loss, which depends on the thermal conductivity of the gas surrounding 
the sensor.  Since different gases, and mixtures, have different thermal conductivities, the indicated pressure 
readings and outputs will also be different.  InstruTech convection gauges (and most other thermal gauges) are 
normally calibrated using nitrogen.  When a gas other than N

2

 is used, correction must be made for the difference 

in thermal conductivity, between N

2

 and the gas in use.  The charts and tables below indicate how different gases 

affect the display and output from an InstruTech convection gauge.    
For N

2

 the calibration shows excellent agreement between indicated and true pressure throughout the range from 

10

-4

 to 1000 Torr.  (As it 

should

, of course, since they're calibrated for N

2

.)  At pressures below 1 Torr, the 

calibration curves for the different gases are similar:  the difference in readings at these low pressures is a 
constant, a function of the difference between thermal conductivities of the gases. 
At pressures above 1 Torr, indicated pressure readings may diverge significantly.  At these higher pressures, 
convection currents in the gauge become the predominant cause of heat loss from the sensor, and calibration 
depends on gauge tube geometry and mounting position, as well as gas properties. 
Generally, air and N

2

 are considered the same as far as thermal conductivity goes, but even these two gases will 

exhibit slight differences in readings at higher pressures.   For example, when venting a system to atmosphere 
using N

2

, you may see readings change by 30 to 40 Torr after the chamber is opened and air gradually displaces 

the N

2

 in the gauge.   

For most other gases, the effect is much more significant, and may result in a hazardous condition, as described 
below. 
 
 
 

Other considerations when using gases other than N

2

 

Flammable or explosive gases

InstruTech CV gauges and modules are not intrinsically safe or explosion proof, and are not intended for the 
measurement, or use in the presence, of flammable or explosive gases or vapors. 
Under normal conditions, the voltages and currents in InstruTech CV Gauges and Modules are too low to cause 
ignition of flammable gases.  However, under certain failure conditions, sufficient energy could be generated to 
cause flammable vapors or gases to ignite or explode.   
Thermal gauges like the InstruTech CV gauges and modules are not recommended for use with flammable or 
explosive gases. 

Moisture / water vapor

In some processes (lyophilization, for example). the gas composition may not change significantly, except for 
moisture content.  Water vapor can significantly change the response of a thermal gauge and correction should 
be made, as you would for any other gas. 

Other contaminants

If your gases condense, coat, or corrode the sensor, the gauge calibration and response to different gases will 
change.   Generally, if the gauge can be "calibrated" ("zero" and "atmosphere" settings), these changes are small 
enough to be ignored.  If you cannot set zero and atmosphere, the gauge should be replaced or cleaned.  (See 
the Appendix C on cleaning for more information.) 

 

 

Содержание CVM211

Страница 1: ...11 S ST TI IN NG GE ER R Convection Vacuum Gauge Module User s Manual Revision 1 05 Sept 25 2007 1475 S Fordham St Longmont CO 80503 Phone 303 651 0551 Fax 303 678 1754 E mail info instrutechinc com w...

Страница 2: ...4 Gases other than N2 5 Flammable Explosive Gases 5 Installation Mechanical 5 Position 5 Vibrations 5 Fittings 5 Electrical 6 Connector Pinouts 6 Setup Operation User Interface Basics 7 Setting up th...

Страница 3: ...Torr to 1 0 mTorr 1 digit LED from 0 9 mTorr to 0 1 mTorr materials exposed to gases gold plated tungsten 304 316 stainless steel glass nickel Teflon internal volume 26 cm3 1 589 in 3 internal surface...

Страница 4: ...parts are specified by Instrutech Inc Unauthorized substitutions or non qualified parts may result in fire electric shock or other hazards To reduce the risk of fire or electric shock do not expose t...

Страница 5: ...ure difference error between your process and the gauge and cause a delay in response to pressure changes Mounting the CVM too close to a gas source will also cause errors in the readings Don t mount...

Страница 6: ...practice is to remove power from any cable prior to connecting or disconnecting it The InstruTech CVM 211 will directly replace GP Mini Convectron modules that have a 9 pin D connector and you can us...

Страница 7: ...to save and go to the next step 6 The readout will display 000 to indicate the unit is in the zero adjust mode To properly set zero the gauge must be evacuated to a pressure below 10 4 Torr Factory de...

Страница 8: ...il due to chemical reaction are generally not salvageable Sensors that fail due to coating may possibly be restored by cleaning See Appendix C for a more complete discussion of contamination and clean...

Страница 9: ...hrough Friday at 303 651 0551 Or e mail us at info instrutechinc com If it becomes necessary to return the unit to InstruTech for service or calibration please call or e mail us for a return authoriza...

Страница 10: ...ceed to next step without changing the zero value press select Readout displays current span value Factory default 760 Torr Is system backfilled with air or N2 to some known pressure above 400 Torr Or...

Страница 11: ...of 0 375 to 5 659 VDC for 0 to 1000 Torr of N2 roughly in the shape of an S curve as shown at right GP adopted the same output curve for most of their Mini Convectron modules and controllers with non...

Страница 12: ...r as thermal conductivity goes but even these two gases will exhibit slight differences in readings at higher pressures For example when venting a system to atmosphere using N2 you may see readings ch...

Страница 13: ...essure is 100 Torr the gauge will read 100 Torr If the gas is Ar when the true pressure is 100 Torr the gauge will read only 10 Torr If you are backfilling your vacuum system with Ar when your system...

Страница 14: ...70 1 3150 1 2910 1 2650 1 0020 1 3920 500 mTorr 1 6833 1 3860 1 5890 1 6640 1 6680 1 1940 1 8260 1 8050 1 9140 1 4690 2 0140 1 Torr 2 2168 1 8180 2 1640 2 1950 2 1720 1 5360 2 2570 2 2470 2 6030 1 976...

Страница 15: ...288 5 233 4 845 5 210 5 220 5 616 5 201 5 522 5 0000 5 699 5 389 6 130 5 697 5 524 5 107 5 389 5 418 7 391 5 719 5 877 10 0000 6 000 5 602 8 041 6 013 5 696 5 250 5 471 5 530 8 041 6 332 6 446 20 0000...

Страница 16: ...auge Module InstruTech Inc page 16 Log Linear Analog Output Voltage vs Pressure 0 1 2 3 4 5 6 7 8 9 0 0001 0 001 0 01 0 1 1 10 100 1000 Pressure Torr Output Signal Volts This is a plot of the formulas...

Страница 17: ...87 359 781 1 Torr 1 00 600 940 970 950 400 1 05 1 04 1 55 745 1 60 2 Torr 2 00 1 14 2 22 1 94 1 71 700 1 62 1 66 4 13 1 59 3 33 5 Torr 5 00 2 45 13 5 4 98 3 34 1 28 2 45 2 62 246 5 24 7 53 10 Torr 10...

Страница 18: ...s steel or ceramic to the gases There is no material that is universally chemical resistant your choice of vacuum gauge as well as all other vacuum components should take into consideration the potent...

Страница 19: ...s such as pump oil However depending on your particular contamination you may need to use a more aggressive solvent Cleaning procedure 1 With the gauge port up pour your cleaning solution into the gau...

Страница 20: ...Furthermore the warranty does not apply to products that have been contaminated or when the product or part is damaged during the warranty period due to causes other than ordinary wear and tear to th...

Отзывы: