![Instron AVE2 Скачать руководство пользователя страница 56](http://html1.mh-extra.com/html/instron/ave2/ave2_operators-manual_2070122056.webp)
Chapter: Set up hardware
56
M16-16905-EN
To examine and, if necessary, change the lens:
1. Remove the lens cover (AVE2 only; there is no lens cover on the SVE2) by gently
pulling it away from the lens. You can now read the focal length of the lens printed
on the body.
2. If you need to change the lens, unscrew the lens from the extensometer and store it
in the case provided.
3. Screw in the new lens and focus it.
4. Replace the lens cover (AVE2 only).
5. If you have changed the lens, recalibrate the extensometer.
AverEdge32™ configuration
The AverEdge32™ configuration changes the distance between the lens and the
specimen compared to the standard configuration. The field of view for each lens is
therefore different.
The values are the same for all supported frame types, both table models and floor
models.
Only the 16 mm lens supports use of the edge-based method for measuring transverse
strain.
The high capacity floor model testers, 5988 and 5989, are not supported in this config-
uration.
Lens focal
length (mm)
Nominal field of view (mm)
Table model testers
(e.g. 5940 and 5960),
ElectroPuls and 8800
Series systems
1
1. when used for dynamic testing on 8800 or ElectroPuls, the 16mm lens is the only available
option.
Floor Model testers
(e.g. 5982, 5984,
5985) and LX static
hydraulic testing
machines
High-capacity floor
model testers
(e.g.5988, 5989)
35
100
130
160
16
(standard)
240
310
390
9
425
560
680
6
620
840
not suitable
Содержание AVE2
Страница 1: ...Instron Video Extensometer AVE2 and SVE2 Operator s Guide M16 16905 EN Revision D...
Страница 8: ...Preliminary Pages 8 M16 16905 EN...
Страница 11: ...11 Preliminary Pages Product Support www instron com Chapter 10 Maintenance 83 General 83 Index 85...
Страница 12: ...Preliminary Pages 12 M16 16905 EN...
Страница 16: ...Chapter Introduction 16 M16 16905 EN...
Страница 24: ...Chapter Specifications 24 M16 16905 EN...
Страница 34: ...Chapter Installation 34 M16 16905 EN Figure 9 AVE2 on a 5940 single column testing machine A...
Страница 46: ...Chapter Installation 46 M16 16905 EN Figure 21 AVE2 on an 8801 series testing machine...
Страница 49: ...49 System interconnections Product Support www instron com Figure 23 Video extensometer connectors 6 1 3 2...
Страница 50: ...Chapter Installation 50 M16 16905 EN...
Страница 76: ...Chapter Set up other software not Bluehill 76 M16 16905 EN...
Страница 84: ...Chapter Maintenance 84 M16 16905 EN...
Страница 87: ......
Страница 88: ...www instron com...