Chapter: Mark specimens
54
M16-16905-EN
•
wicking can occur when testing woven specimens and the wet ink is drawn, by
capillary action, into the fibers of the specimen.
If bleeding or wicking is a consistent problem, you should consider using tape dots.
•
Loss of detection of ink dots during tests can be caused by:
•
change of specimen color during testing so that contrast between the dot and
the specimen is lost - consider using a different ink color
•
fading of the dot due to excessive extension during the test - consider using tape
dots
•
flaking of the ink from the specimen - make sure the ink is compatible with the
specimen material and consider testing the specimen while the ink is still wet
•
the dot moves outside the field of view of the lens because specimen extension
was larger than expected - use a lens with a larger field of view
Содержание AVE2
Страница 1: ...Instron Video Extensometer AVE2 and SVE2 Operator s Guide M16 16905 EN Revision D...
Страница 8: ...Preliminary Pages 8 M16 16905 EN...
Страница 11: ...11 Preliminary Pages Product Support www instron com Chapter 10 Maintenance 83 General 83 Index 85...
Страница 12: ...Preliminary Pages 12 M16 16905 EN...
Страница 16: ...Chapter Introduction 16 M16 16905 EN...
Страница 24: ...Chapter Specifications 24 M16 16905 EN...
Страница 34: ...Chapter Installation 34 M16 16905 EN Figure 9 AVE2 on a 5940 single column testing machine A...
Страница 46: ...Chapter Installation 46 M16 16905 EN Figure 21 AVE2 on an 8801 series testing machine...
Страница 49: ...49 System interconnections Product Support www instron com Figure 23 Video extensometer connectors 6 1 3 2...
Страница 50: ...Chapter Installation 50 M16 16905 EN...
Страница 76: ...Chapter Set up other software not Bluehill 76 M16 16905 EN...
Страница 84: ...Chapter Maintenance 84 M16 16905 EN...
Страница 87: ......
Страница 88: ...www instron com...