Quick start guide |
Safety Information
09
Precautions
Danger of EXPLOSION:
Do not use the Eazy PRO in the presence of flammable
anesthetics.
Risk of ELECTRICAL SHOCK:
Never open the equipment. When necessary, this
must be done by authorized individuals.
Do not use the equipment in the presence of magnetic resonance devices.
This equipment was designed to be resistant to electromagnetic interference.
However, equipment performance can be affected if in the presence of strong
sources of electromagnetic interference or radio frequencies, such as mobile
phones, radio communicators, etc.
Electromagnetic Compatibility
The Eazy PRO is a device built according to NBR and IEC standards and therefore
is completely safe for the patient and the user. However, all safety precautions
described below must be followed
The operation of the Eazy PRO may be affected by the presence of
electromagnetic power supplies, such as electrosurgical equipment and
computer tomography (CT).
WARNING:
Using the Eazy PRO requires special precautions concerning
Electromagnetic Compatibility in compliance with the information contained in this
manual.
Mobile and portable RF communications equipment, such as a cellphones, may
affect the functioning of the Eazy PRO.
Maximum length of accessory cables in order to comply with the requirements of
Electromagnetic Compatibility is 2,5 m.
All parts and accessories, listed below, follow the requirements for Electromagnetic
compatibility.
• Pair of disposable adhesive pads, adult size
• Pair of disposable adhesive pads, child size (optional)
• Power supply for charging internal battery
• USB cable
WARNING:
using cables and accessories different from the ones specified above,
except for cables and accessories sold by Instramed as replacement pieces, may
result in emission gain or immunity decrease of the equipment.
The Eazy PRO must not be used too close to or piled over other equipment.
More information on this topic is available in the Eazy PRO User Manual
present on the CD which accompanies the product.