Quick start guide |
Safety Information
08
Warnings
IMPORTANT:
this equipment may only be operated by qualified technical
personnel. Read this guide carefully before using the equipment.
WARNING: THE PATIENT MUST BE PLACED ON NON CONDUCTIVE
SURFACES. DO NOT USE WET OR METALLIC SURFACES AND, IF
NECESSARY, DRY THE CHEST BEFORE APPLYING THE SHOCK.
WARNING: DO NOT TOUCH THE PATIENT, THE EQUIPMENT, THE
ACCESSORIES NOR ANY METALLIC OR CONDUCTIVE SURFACE WHICH IS
IN CONTACT WITH THE PATIENT DURING THE DEFIBRILLATION.
WARNING: DO NOT CONNECT THE PATIENT TO THE EAZY PRO WHEN THE
EQUIPMENT IS CONNECTED TO THE ELECTRIC CURRENT.
WARNING: THE PATIENT MUST BE COMPLETELY STILL DURING THE
CARDIAC RHYTHM ANALYSIS PHASE. DO NOT GIVE CARDIAC MASSAGE AT
THIS POINT.
WARNING:
risk of explosion if the equipment is operated in the presence of
flammable liquids or gases.
WARNING:
always check the general state of the equipment, the battery and the
accessories before using it.
NOTICE:
each and every repair to the equipment can only be done by Instramed’s
authorized technical assistance centers.
NOTICE:
the use of the Eazy PRO is restricted to one patient at a time.
NOTICE:
the applied parts are protected against defibrillation discharge; during
discharge there may be baseline variation.
NOTICE:
avoid connecting the patient to several items of equipment at the same
time. The limits of current leakage may be exceeded.
NOTICE:
the applied parts intended to come into contact with the patient have been
evaluated and comply with the directives and principles of ISO 10993-1.
NOTICE:
when removing the equipment from the package, carefully verify if there
is any abnormality or visible damage in the device or its accessories, caused by
impact or mishandling during transportation. In case of irregularities, please contact
Instramed.
NOTICE:
disposable accessories and any other components must be disposed of
according to the norms of hospital waste disposal.
Adverse effects
Superficial burns may occur to the patient’s skin on the area in contact with the
electrodes. To minimize this effect, apply the pads soon after removing their
protective envelope and press them firmly to the patient’s skin.
The skin should be dry, or electrical current leakage may happen, enlarging burn
area and reducing treatment efficiency.