IfU TECOTHERM NEO TN300 EN - 01.doc
Page 66 of 76
Ref No. TN300 EN-01
11.4 Alarm Fluid level too low
see Section 11.6, too
Fluid level too low. Little or no Fluid in the container.
NOTE
When alarm is appearing push key T5 to silence audible
alarm (AUDIO paused,
is appearing at the display.)
Caused by
Leakage in the TECOTHERM NEO device / Hydraulic Module.
Lack of fluid caused by leakage in hoses, at couplings or in mattress
In the container
Fluid volume is less than 1/3 for at least 10 seconds.
There are now 8 minutes to resolve the problem.
Attention
Check immediately whether liquid is escaping or
was escaping from the device.
Check immediately whether a leakage is in the mattress or hoses.
If a big leak or fluid volume is seen below or near the device
immediately shut off
the
device. Proceed as in other alarm cases. Follow the instructions in the Main MENU.
Contact your local service.
Elimination/ Management
If leakage or / and indication of escaped liquid/ traces of liquid in your urgent
immediate inspection was not found refill the system. For refilling TECOmed follow the
procedure of section 8.5.
Continue treatment following Main MENU instructions.
If alarm persists and symbol is not disappearing you have to make troubleshooting
more systematically.
Case 1
Check / inspection of the device
Inspect the support area below the device, device bottom and lower casing parts for
traces and indications of liquid/ moisture.
If you find a small or medium leak: Immediately Shut Off the device following the
instructions in the Main MENU. Unplug mains. Contact / inform your local service.
If you find a small leak contact your service for assistance how to proceed.