9.4 NIBP Alarm Information
The following situations may cause a longer measurement time or
unreliable values:
• Animal Movement
• Severe Shock
• Low Heart Rate
• Arrhythmia
• Rapid Pressure Changes
• Extremely Large Animals
Error
Cause
SysErr
Self-test fail
SysErr2
NIBP module system error
CuffLoose
Cuff is too loose or cuff not connected
CuffErr
Using small cuff in big cuff mode
Leakage
Valve or gas circuit leak
PressErr
NIBP Valve is not working appropriately
Weak
Animal’s pulse is too weak or cuff is loose
OveRange
Animal’s blood pressure exceeds the measurement range
Movement
During measurement, interruption in signal or too much
interference
Protect0
Cuff pressure exceeds the range 300mmHg
Saturate
Large signal amplitude caused by interference
TimeOut
Big Cuff: cuff pressure over 2kPa(15mmHg) lasting for more
than 3 minutes.
Small Cuff: cuff pressure over 0.67kPa (5mmHg) lasting for
more than 90s
Reset
NIBP module reset