The pacing signal was not greater than the height of R wave.
•
The height of T wave is less than 1/3 of that of R wave.
•
P wave should be much smaller than T wave.
•
In order to obtain 1 MV calibrated ECG wave, ECG calibration should be
performed. At this time,
the screen will prompt that ‘animals cannot be monitored
during calibration’.
Fig. 6-3 Standard ECG waveform
Using Five Lead ECG Device
Users can arrange the leads on channel 1, channel 2 and channel 3 according to
their own needs. The lead labels on the three channels are displayed above the
corresponding waveforms and can be changed in the ECG menu. For channel 1,
channel 2 and channel 3, select the appropriate lead from I, II, III, AVR, AVL, AVF
and V.
Warnings
•
If the electrode is correct and the ECG waveform is not accurate, replace the
lead.
•
Interference from ungrounded instruments near animals and ESU interference
may cause waveform problems.