User Manual
28 / 33
animated balloon
is not moving
Switch the device on and off again to reset
Test Starts
before you start
blowing
Vigorous handling of
the device
Keep the device as stable as possible after
starting a test
Device
disconnected
during test
The device is turned off
accidentally or due to
rough handling during
use
Switch the device on again and proceed with a
new test
BLE connection
dropped
Reconnect the device and proceed with a new
test
Test quality grade
always low
Not performing test
correctly
Repeat the test following the rules and
conditions specified in the Performing a Lung
Function Test section of this user manual.
Measurement
error screen
showed up
Flow limit exceeded
This device is intended to measure 0-14 L/s.
Spiroway mouthpiece is
dirty
Replace Spiroway
Spiroway mouthpiece is
damaged
Replace Spiroway
Device malfunction
Contact manufacturer
Device error
screen showed
up
Spiroway mouthpiece is
installed incorrectly
Refer to user manual for proper installation of
Spiroway
There is a foreign object
between the sensors
Check Spiroway to ensure that the lumen is
not obstructed
Spiroway mouthpiece is
dirty
Replace Spiroway
Spiroway mouthpiece is
damaged
Replace Spiroway
Device is not
responding to
button
Device error
Remove the batteries, wait 30 seconds and
reinstall batteries
For any other technical queries please call our customer service on +90 312 988 03 08 or
e-mail [email protected]
First Pub. Date: 24.09.2018