16 / 33
7. The test results will be displayed on the app screen. If the test fails, give feedback
and guide the patient for another trial. Encourage them to breathe deep and fast and
try to reach at least 12 seconds.
➢
The Slow Vital Capacity (SVC) Test Breathing Maneuver:
1. Ensure that the device is connected. Select the SVC test mode and the test screen
will appear.
2. Enter the required ambient conditions (makes sure you entered the correct values as
the measurement may be significantly affected by a wrong value) like temperature
and relative humidity and then adjust zero flow level for the device.
3. Tell the patient to wear a nose clip and ask the patient to place the mouthpiece in
his/her mouth, past his/her teeth and ensure that his/her lips are tightly sealed
around the mouthpiece.
4. When the test starts, the patient should inhale and exhale normally at least 4 times,
then the patient should inhale as deep as the patient can and fill his/her lungs
completely.
5. After that, the patient should exhale the whole air in his/her lungs gently and slowly
until the patient feels that all the air in his/her lungs feels completely empty without
breaking the seal of his/her lips.
6. When performing this breathing maneuver, the patient must make sure to keep
blowing until the patient feels like the patient has completely emptied his/her lungs.
7. The test can also be performed by performing the breath maneuver in the opposite
direction. When the test starts, the patient should inhale and exhale normally at least
4 times, then the patient should exhale as deep as the patient can and empty his/her
lungs completely. After that, the patient should inhale all the air in his/her lungs until
s/he feels completely full without breaking the seal of his/her lips.
8. The patient may remove the mouthpiece from his/her mouth and resume normal
breathing once the breathing maneuver is complete.
9. The test results will be displayed on the app screen. Give feedback to the patient on
his/her effort by looking at the test results. The patient will need to perform 2 more
tests by repeating this breathing maneuver. However, please make sure that the
patient has time to rest between tests and feels well enough to continue.
3.4.3.
End of the Tests
Once all tests have been satisfactorily completed, you will be able to view the session results
on the results page of the app. After the end of the spirometry session, remove and
immediately dispose of the mouthpiece by pushing the notch without touching the top part of
the mouthpiece. Turn the device off by pressing the power button, close the cap and store
the device according to the storage requirements until next use.