Guidance and manufacturer´s declaration – electromagnetic immunity
Cervella
is intended for use in the electromagnetic environment specified below. The customer or the user
of the
Cervella
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment
- guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
6 V in ISM and amateur
radio bands between
0,15 MHz and 80 MHz
10 V/m
80 MHz to 2.7 GHz
385MHz-5785MHz Test
specifications for
ENCLOSURE PORT
IMMUNITY to RF
wireless communication
equipment (Refer to
table 9 of IEC 60601-1-
2:2014)
3V
150 kHz to 80 MHz
6 V in ISM and amateur
radio bands between
0,15 MHz and 80 MHz
10 V/m
80 MHz to 2.7 GHz
385MHz-5785MHz Test
specifications for
ENCLOSURE PORT
IMMUNITY to RF
wireless communication
equipment (Refer to
table 9 of IEC 60601-1-
2:2014)
Portable and mobile RF
communications equipment
should be used no closer to any
part of the
Cervella
, including
cables, than the recommended
separation distance calculated
from the equation applicable
to the frequency of the
transmitter.
Recommended separation
distance
P
V
d
]
5
.
3
[
1
=
P
V
d
]
12
[
2
=
P
E
d
]
5
.
3
[
1
=
80 MHz to
800 MHz
P
E
d
]
7
[
1
=
800 MHz to
2.7 GHz
where
p
is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and
d
is the
recommended separation
distance in metres (m).
b
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site