TOP-X100NR USER MANUAL
Innomed Medical Inc.
36
Burst – Digital storage of 10 images.
Rapid – Continuous digital image storage. In this and in the previous mode,
the image creation frequency appears next to the symbol.
The following line contains the accumulated fluoroscopy time.
To the left in the bottom row is the magnification factor. To the right is the symbol
indicating ABS active state. In the center are the different fluoroscopic mode states:
Continuous fluoroscopy
Impulse fluoroscopy. In this operating mode the impulse frequency appears
under the symbol.
8.5.2 Setting fluoroscopy x-ray voltage (kV)
You can set this value on the control panel with the soft buttons next to the fluoro kV.
Use + (13) button next to the symbol to increase the value, and – (14) to decrease it.
Continuously holding the buttons changes the value within the allowed limits in 1kV
steps.
8.5.3 Setting fluoroscopy x-ray current (mA)
You can set this value on the control panel with the soft buttons next to the fluoro
mA. Use + (19) button next to the symbol to increase the value, and – (20) to
decrease it. Continuously holding the buttons changes the value within the allowed
limits in 0.1mA steps
8.5.4 Fluoroscopy time
The clock on the display indicates fluoroscopy time, and only operates when x-ray
emits from the tube. The timer can be reset with the soft button (8) next to the timer
reset symbol. When exceeding the “beep time” value set in the install program during
installation, the control panel warns with a continuous sound that fluoroscopy
duration is nearing what is dangerous to patient health. When reaching the allowed
maximum fluoroscopy time (also set in the install program), the generator
automatically disables the x-ray. The examination can continue only when
fluoroscopy timer is reset.
8.5.5 Zoom
You can access the image intensifier’s zoom functions with this parameter. You can
select among normal and max. four magnified image displays. During magnification
fluoroscopy current is automatically multiplied. To do this, enter fluoroscopy
parameter setting with the “More options” soft button (12).
The following screen appears: