Recommended separation distances between portable and mobile RF
telecommunication equipment and TOP-X generator
TOP-X generator is intended for use in electromagnetic environment where
electromagnetic disturbances are under control. Customer and user of TOP-X
generator can help on preventing electromagnetic influence between portable and
mobile RF telecommunication equipment (transmitters) and TOP-X generator by
determining minimum separation distance between TOP-X generator and
telecommunication equipment that calculated from maximum output power of
transmitter, the following way:
Separation distance as a function of transmitter frequency
m
Maximum
output power of
transmitter
W
150 kHz – 80 MHz
d =
P
⋅
2
.
1
80 MHz – 800 MHz
d =
P
⋅
2
.
1
800 MHz – 2.5 GHz
d =
P
⋅
3
.
2
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
If the transmitter power is not in the table, protective distance
d
in meter (m) can be
calculated by formula depending on frequency, where
P
is maximum transmit power
supplied by manufacturer of the transmitter in watts (W).
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, object and people.