Innomed Medical
Heart Screen 60G-80G-L
Service manual
2014.01.09 26/84
D-5450-1-B_HS60G80GL_SM_ENG.doc
6.5.3 CMRR check
For CMRR setting connect the patient cable and short all the electrodes except for the “N” point. Apply a 1Vpp +/-
0.5% amplitude 50Hz sine signal between the amplifier’s ground point and the shorted electrodes. Set the device to the
highest, 0,5mV/cm sensitivity (x2) and 25 mm/sec speed. With the channel selector monitor all the channels one-by-
one.
During monitoring measure the signal on the AD input (AD1.5 pin) with a scope by synchronizing it to the CH3MPX
(U20.15) signal. The eight channels’ phases must be visible on the scope. The individual phase’s time is 125
R
sec.
During continuous monitoring try to decrease the error signal with P6-and resistor R176 so approximately the same
error signal is generated on all the channels. Watch the error signal on the scope.
The CMRR is acceptable is the measured amplitude is less than 250
R
Vpp, or 74 dB.
P1 P2
Measured
lead
N
all other
I
N
all other
II
N
all other
V1
N
all other
V2
N
all other
V3
N
all other
V4
N
all other
V5
N
all other
V6
6.5.4 Checking blocking
For the measurement connect the patient cable and short all the electrodes with the exception of the “R” lead. Apply
+320 mV DC voltage between the “R” lead and the other leads. Set the device to highest, 0.5mV/cm sensitivity (x2)
and 25 mm/sec speed.
Monitor all the channels one-by-one during continuous monitoring. During monitoring give a blocking command.
After blocking the signal must remain within 100
R
V. Allow the signal to run for at least 10 seconds on all the
channels after blocking, the baseline must remain in the +/-100
R
V range all along.
Repeat the measurement with -320 mV DC as well.
6.5.5 Checking frequency characteristic
Apply (16000 +/-0.5%
R
Vpp / 150Hz sine) + DC signal on the amplifier input according to the tables from a ground-
independent generator. DC should be +320, -320 mV. Set the device to 4mV/cm sensitivity (x1/4) and 25 mm/sec
speed.
Connect the patient cable according to the following tables, start the individual channels with the channel selector and
measure the signal amplitude.
a/
In case of DC = -320 mV
The signal’s amplitude on the measured lead should be at least 11200
R
V.
P1
P2
Measured lead
L
all other
I
R
all other
II
V1 all
other V1
V2 all
other V2
V3 all
other V3
V4 all
other V4
V5 all
other V5
V6 all
other V6
Содержание Heart Screen 60 G
Страница 51: ...Innomed Medical Heart Screen 60G 80G L Service manual 2014 01 09 51 84 D 5450 1 B_HS60G80GL_SM_ENG doc BOTTOM ...
Страница 55: ...Innomed Medical Heart Screen 60G 80G L Service manual 2014 01 09 55 84 D 5450 1 B_HS60G80GL_SM_ENG doc ...
Страница 56: ...Innomed Medical Heart Screen 60G 80G L Service manual 2014 01 09 56 84 D 5450 1 B_HS60G80GL_SM_ENG doc ...
Страница 57: ...Innomed Medical Heart Screen 60G 80G L Service manual 2014 01 09 57 84 D 5450 1 B_HS60G80GL_SM_ENG doc ...
Страница 58: ...Innomed Medical Heart Screen 60G 80G L Service manual 2014 01 09 58 84 D 5450 1 B_HS60G80GL_SM_ENG doc ...
Страница 59: ...Innomed Medical Heart Screen 60G 80G L Service manual 2014 01 09 59 84 D 5450 1 B_HS60G80GL_SM_ENG doc ...
Страница 60: ...Innomed Medical Heart Screen 60G 80G L Service manual 2014 01 09 60 84 D 5450 1 B_HS60G80GL_SM_ENG doc ...
Страница 61: ...Innomed Medical Heart Screen 60G 80G L Service manual 2014 01 09 61 84 D 5450 1 B_HS60G80GL_SM_ENG doc ...
Страница 62: ...Innomed Medical Heart Screen 60G 80G L Service manual 2014 01 09 62 84 D 5450 1 B_HS60G80GL_SM_ENG doc ...
Страница 64: ...Innomed Medical Heart Screen 60G 80G L Service manual 2014 01 09 64 84 D 5450 1 B_HS60G80GL_SM_ENG doc ...
Страница 66: ...Innomed Medical Heart Screen 60G 80G L Service manual 2014 01 09 66 84 D 5450 1 B_HS60G80GL_SM_ENG doc ...
Страница 69: ...Innomed Medical Heart Screen 60G 80G L Service manual 2014 01 09 69 84 D 5450 1 B_HS60G80GL_SM_ENG doc Bottom ...
Страница 70: ...Innomed Medical Heart Screen 60G 80G L Service manual 2014 01 09 70 84 D 5450 1 B_HS60G80GL_SM_ENG doc ...
Страница 71: ...Innomed Medical Heart Screen 60G 80G L Service manual 2014 01 09 71 84 D 5450 1 B_HS60G80GL_SM_ENG doc ...
Страница 72: ...Innomed Medical Heart Screen 60G 80G L Service manual 2014 01 09 72 84 D 5450 1 B_HS60G80GL_SM_ENG doc ...
Страница 73: ...Innomed Medical Heart Screen 60G 80G L Service manual 2014 01 09 73 84 D 5450 1 B_HS60G80GL_SM_ENG doc ...
Страница 74: ...Innomed Medical Heart Screen 60G 80G L Service manual 2014 01 09 74 84 D 5450 1 B_HS60G80GL_SM_ENG doc ...
Страница 75: ...Innomed Medical Heart Screen 60G 80G L Service manual 2014 01 09 75 84 D 5450 1 B_HS60G80GL_SM_ENG doc ...
Страница 76: ...Innomed Medical Heart Screen 60G 80G L Service manual 2014 01 09 76 84 D 5450 1 B_HS60G80GL_SM_ENG doc ...
Страница 77: ...Innomed Medical Heart Screen 60G 80G L Service manual 2014 01 09 77 84 D 5450 1 B_HS60G80GL_SM_ENG doc ...
Страница 78: ...Innomed Medical Heart Screen 60G 80G L Service manual 2014 01 09 78 84 D 5450 1 B_HS60G80GL_SM_ENG doc ...
Страница 80: ...Innomed Medical Heart Screen 60G 80G L Service manual 2014 01 09 80 84 D 5450 1 B_HS60G80GL_SM_ENG doc ...
Страница 81: ...Innomed Medical Heart Screen 60G 80G L Service manual 2014 01 09 81 84 D 5450 1 B_HS60G80GL_SM_ENG doc ...
Страница 83: ...Innomed Medical Heart Screen 60G 80G L Service manual 2014 01 09 83 84 D 5450 1 B_HS60G80GL_SM_ENG doc ...
Страница 84: ...Innomed Medical Heart Screen 60G 80G L Service manual 2014 01 09 84 84 D 5450 1 B_HS60G80GL_SM_ENG doc ...