34
VIEW 5 Pro
730
VIEW
1. Press the Current key;
2. Select the current range;
3. Confirm.
After pressing the Current key, the range options are displayed in Display Bar C and can be switched with navigation (up
and down) keys. In the process of setting the current range, the "A" function indicator light on the right side of Display Bar
C is on.
Fixed current range
When the peak factor is set to 3, the optional current range includes 0.5A, 1A, 2A, 5A, 10A and 20A.
When the peak factor is set to 6, the optional current range includes 0.25A, 0.5A, 1A, 2.5A, 5A and 10A.
Automatic current range
The instrument supports automatic current range, and the automatic range options are consistent with the fixed range
options. The operation method is the same as that of above fixed range settings. When switching in Display Bar C, select
"Auto" option and press "Enter" key to confirm. At this time, the current has been set to automatic range mode, and the
current "Auto" indicator light on the panel is on.
Wiring mode
Power range calculation
1P2W
Voltage range*current range
1P3W
Voltage range*current range*2
3P3W
3V3A
3P4W
Voltage range*current range*3
4.3 Power Range
The measuring range of active power (power range) is determined by the wiring mode, voltage range and current range.
The details are as follows.
⚠ CAUTION
When the result of voltage range × current range reaches or exceeds 1,000W, the display unit becomes "kW". When it
exceeds 1,000kW, the unit becomes "MW".
4.4 Setting of the Range for External Current Sensor
Operating steps