![INNO Instrument VIEW 730 Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/inno-instrument/view-730/view-730_user-manual_2062314020.webp)
20
VIEW 5 Pro
730
VIEW
1. Press the Setup key;
2. Select the "u.rate" menu;
3. Confirm;
4. Set update rate options;
5. Confirm;
6. Exit and close the setting menu.
It is operated according to the above steps to enter the data update rate setting interface. The update rate option is
located in Display Bar C. The update rate options can be switched through the navigation (up and down) key, and the
"Enter" key is pressed to complete the setting. The instrument returns to the superior menu to display, and pressing the
"Hold/ESC" key can exit the setting interface.
Description
The data update rate options supported by the instrument are 0.1 s, 0.25 s, 0.5 s, 1 s, 2 s, 5 s and Auto. After the update
rate is set, the "Update" indicator light on the panel will flash according to the set update period. When the "Auto" option
is selected, the instrument automatically selects the matching update rate according to the frequency of the input signal,
and the "Update" indicator light flashes according to the self-matching update period. When the frequency of the input
signal cannot be detected, the automatic update rate is updated by default according to the time-out period.
3.4 Data Update Rate Setting
Operating steps
1. Press Utility key;
2. Select the "reso" menu;
3. Confirm;
4. Set the number of display digits;
5. Confirm;
6. Exit and close the setting menu.
According to the above steps, the switch option "4" or "5" is pressed through navigation (up and down) in the Display Bar
C, and "Enter" key is pressed to confirm and complete the setting. The instrument returns to the superior menu to display,
and pressing the "Hold/ESC" key can exit the setting interface.
Description
If the number of display digits is set to 4, the data will be displayed up to 4 digits; if the number of display digits is set to 5,
the data will be displayed up to 5 digits. If the measured value is 100V, and the number of digits is 4, the displayed value
is 100.0 V; when the number of digits is 5, the display value is 100.00 V. The integral value and time are not affected by the
display digit setting.