ROUTINE MAINTENANCE AND SERVICE 

 

 

At the latest after six months of operation, replace the contents of the bath. Periodically check 

the cleanliness and filling of bath. During sanitizing the water temperature must not be higher than 
40°C.  Cleaning of all external surfaces of the instrument should be done with a soft cloth. You can 
use  common  cleansers.  Do  not  leave  the  instrument  on  a  surface  contaminated  from  corrosive 
substances. This could cause damage of paint on the instrument. The surface of the instrument clean 
with a soft cloth.  

Cleaning the condenser perform at least 1x per month by sweeping, or air pressure, or CO

2

Fan  cover by pulling upward pull of metal trails. Unscrew the fastening screws from the  cooler on 
the condenser side.  Loose cover first pull up upwards 3 cm, than pull to the side.  Loosen the other 
side  of  the  cover  so  that  you  can  pull  the  whole  cover  of  metal  trails.  Now  you  can  clean  the 
condenser. 
Warning:  When cleaning, unplug the instrument! 
 
 

 

 

 

                       

Ejecting the fan                              Putting the fan in place            Unscrewing the cover 

 

 

 

 

 

 

                      

Removing the cover                Condenser ready for cleaning 

 

5. 

ACCESSORIES 

 
2 pcs   Set of hoses and couplings for connection RVO 400 

 

 

 

 

Содержание COR 400

Страница 1: ...rvice INGOS Ltd K Nouzovu 2090 143 00 Prague 4 Czech Republic Tel 420 225 983 400 420 225 983 409 Fax 420 244 403 051 e mail instruments ingos cz www ingos cz 17 7 2019 CIRCULATION CHILLER COR 400 Use...

Страница 2: ...Chiller Warning Using devices other than as described in the instructions can cause security breaches 2 DESCRIPTION AND SPECIFICATION 2 1 Description The instrument consists of a box in which is therm...

Страница 3: ...ct all packing material If any item is missing contact the manufacturer or your supplier Place the instrument on the table and familiarize yourself with the distribution and function of the controls a...

Страница 4: ...e coupling it is necessary to moisten the end 3 2 Temperature controller Display along with a button keyboard provide user comfort for direct control The function of the individual keys is as follows...

Страница 5: ...or in 30 seconds the thermostat when there is no button operation will return to normal operating mode Setting item Range setting parameters Factory setting code Temperature difference 1 C 16 C 1 C F0...

Страница 6: ...f paint on the instrument The surface of the instrument clean with a soft cloth Cleaning the condenser perform at least 1x per month by sweeping or air pressure or CO2 Fan cover by pulling upward pull...

Страница 7: ...be performed by the manufacturer or its authorized body Warning For using water as coolant set in mode F2 lower limit of the temperature at 1 C and above 7 WASTE DISPOSAL At the end of life discard th...

Отзывы: