4.
ROUTINE MAINTENANCE AND SERVICE
At the latest after six months of operation, replace the contents of the bath. Periodically check
the cleanliness and filling of bath. During sanitizing the water temperature must not be higher than
40°C. Cleaning of all external surfaces of the instrument should be done with a soft cloth. You can
use common cleansers. Do not leave the instrument on a surface contaminated from corrosive
substances. This could cause damage of paint on the instrument. The surface of the instrument clean
with a soft cloth.
Cleaning the condenser perform at least 1x per month by sweeping, or air pressure, or CO
2
.
Fan cover by pulling upward pull of metal trails. Unscrew the fastening screws from the cooler on
the condenser side. Loose cover first pull up upwards 3 cm, than pull to the side. Loosen the other
side of the cover so that you can pull the whole cover of metal trails. Now you can clean the
condenser.
Warning: When cleaning, unplug the instrument!
Ejecting the fan Putting the fan in place Unscrewing the cover
Removing the cover Condenser ready for cleaning
5.
ACCESSORIES
2 pcs Set of hoses and couplings for connection RVO 400