INGOS Ltd. COR400 17.7.2019
Strana 3
2.2 Specification
Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1/4 HP
Performance 22°C =>7°C . . . . . . . . . . . . . . . . . . . . . . . . 80 l/hour
Minimum temperature . . . . . . . . . . . . . . . . . . . . . . . . . . 1°C for water
Minimum temperature . . . . . . . . . . . . . . . . . . . . . . . . . . -20°C for antifreezing mix
Regulation accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . .±1°C
Volume of cooling tank . . . . . . . . . . . . . . . . . . . . . . . . 20 l
Displacement pump . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 m
Pump flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120 l/hour
Cooling gas. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . R 290
Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 230V, 50Hz
Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . max. 391 VA
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 kg
Dimensions (w x d x h) . . . . . . . . . . . . . . . . . . . . . . . . . 530 x 330 x 474 mm
The instrument is designed for an ambient temperature 15 - 35 °C; with humidity up to 80%
without acid and caustic vapors.
3.
COMMISSIONING
3.1 Unpacking
Carefully unpack the instrument from the shipping container. Check the exterior of the
instrument, suffer during the transport from visible damage. Should any damage occur during the
transport, contact the shipping company.
According to the delivery note check all items. In case of discrepancy carefully inspect all
packing material. If any item is missing, contact the manufacturer or your supplier.
Place the instrument on the table and familiarize yourself with the distribution and function of
the controls and connections. Before operating read the instructions of this manual thoroughly.
Place the chiller so that the heat generated by the chiller may be sufficiently ventilated,
must not be placed near radiant heat or direct sunlight. The chiller should be placed on a
horizontal surface. Do not place objects the chiller that would prevent air circulation.
The chiller can in no case be put on its side even during transport!
3.1.1 What you will need
To put into operation, you will need the following items:
1.
Tubing with insulation and couplings for connection to a rotary vacuum evaporator cooler.
2.
20 - 25 liters of antifreeze with a freezing point of about 5°C lower than the lowest temperature
set on the thermostat. When setting the lowest limit of temperature of 1°C and above, you can be
used water.
Fill the plastic tank with liquid to the overflow.