Continuous Gas Monitoring Analyzer 

 

Содержание IR-8400D

Страница 1: ...IR 8400D Continuous Gas Monitoring Analyzer ...

Страница 2: ......

Страница 3: ...2 Operator s Manual Infrared Industries 25590 Seaboard Lane Hayward CA 94545 Phone 800 344 0321 Fax 510 782 8101 www infraredindustries com ...

Страница 4: ...3 ...

Страница 5: ...AS CONNECTIONS 17 CONFIGURING THE ANALYZER 18 STANDBY 18 IDENTIFYING ACTIONS WHEN PUMP IS On 19 Menu 1 20 MENU 2 21 CALIBRATION AND OPERATION 22 GENERAL OPERATION 22 MEASURING GASES 22 CALIBRATION 22 RECOMMENDED CALIBRATION SCHEDULE 23 Auto Calibration 23 IR SAMPLE CONDITIONER 28 IR1150 1151 SAMPLE CONDITIONER 28 INSTALLATION 30 INITIAL SETUP 30 OPERATION 30 FILTER REPLACEMENT 30 WARRANTY 32 Warra...

Страница 6: ...5 ...

Страница 7: ...cables and connectors for proper mating to the PC Boards Check the gas sample lines from the bulkhead fittings to the optical bench for integrity If all internal components look proper replace the display board and screw the front window securely in place Remove the junction box cover Inspect the fuse and terminal board The cover can be left off for installation purposes R E P O R T I N G D A M A ...

Страница 8: ...7 ...

Страница 9: ...rocessed and used to derive a direct measurement of the measured gas concentration in the sample gas Figure 1 IR8400D The Infrared Industries 8400D series models are non dispersive infrared NDIR gas analyzers capable of continuously monitoring various gases in a sample stream for extended periods of time The IR 8400D is designed specifically for the measurement of Hydrocarbons Carbon Dioxide Carbo...

Страница 10: ...ther environments the analyzers provide as standard an LCD display digital RS232 output analog output and an isolated linear 4 20 mA current output Features include a solid state detector a direct reading display a low energy infrared source a sample cell that is easy to remove and clean and many field serviceable components ...

Страница 11: ... than 0 5 F S Output response time 0 90 Less than 10 seconds electronic Power required Standard 100 240 VAC maximum 117 Vrms 10 50 60 Hz Maximum draw 2 5 Amps Optional 12 VDC 24 VDC Warm up time for specified performances Less than 1 hour Sample flow rate 1 scfh 1 2 sipm recommended 2 scfh 1 sipm maximum exhaust to ambient Environmental temperature range 40 to 122 F 40 to 50 C Materials of constru...

Страница 12: ...11 ...

Страница 13: ...E N T Oscilloscope Tektronix Model 912B or equivalent Digital Voltmeter Fluke Model 8000A or equivalent Z E R O A N D C A L I B R A T I O N G A S E S Nitrogen in a pressurized cylinder zero gas Calibration gas near full scale concentration with a nitrogen balance in a pressurized certified cylinder G A S H A N D L I N G E Q U I P M E N T Pressure regulators for zero and span gas cylinders Teflon T...

Страница 14: ...13 ...

Страница 15: ...Internal temperatures over 140 F 60 C must be avoided A sunshade as discussed later should be utilized Line voltage surge protectors are provided Infrared Industries may include additional protection devices at the time of installation if deemed necessary M E C H A N I C A L This analyzer is designed to be mounted vertically on a wall or panel see Figure 2 IR 8400D Series Outline Dimensions You mu...

Страница 16: ...15 Figure 2 IR 8400D Outlying Dimensions ...

Страница 17: ... device must be included in the building installation This device should be in close proximity to the equipment and within easy reach of the operator A single phase power source is required to operate the instrument The standard instrument requires 100 to 240 volts AC The hot neutral and ground leads should be connected to the appropriate line power inputs using the insulated terminals provided wi...

Страница 18: ... a feed of clean and dry sample gas at atmospheric pressure for analysis The sample flow rate should not exceed 2 SCFH 1 SIMP Since the analyzer is usually calibrated with the sample cell at atmospheric pressure pressurized gases must be reduced ahead of the analyzer A typical gas handling system would include a pressure regulator a coalescing particulate filter a multiple inlet valve for introduc...

Страница 19: ...you are not sure which button to use S T A N D B Y The 8400D will return to this screen after exiting the menu selection It is possible to customize the display screen settings by going to Display Settings in Menu 2 This allows the operator to select one or as many as all of the values to be displayed on the screen M E A S U R E Allows manual on off operation of the pump This function can also be ...

Страница 20: ...ENU to MEASURE PRINT HOLD and RECORD Pressing the MEASURE key once more while the pump is running will stop the pump and return you to the standby menu P R I N T Prints data to a printer connected to the serial port located in the Junction Box H O L D R E S U M E Selecting the HOLD key will lock the display values to capture that moment in time Once the HOLD key has been initiated the HOLD enuncia...

Страница 21: ... Any gas you do not wish to change the calibration on should be set to which indicates no value change O2 is self calibrating on atmospheric O2 air To enter the tag values of your calibration gas use the GAS key to select the gas Press to change the value of the highlighted numeral and select the NEXT key to advance to the following numeral When finished press Continue to advance to the next seque...

Страница 22: ...displayed S Y S T E M I N F O R M A T I O N This screen contains information about your system PEF propane equivalency factor the serial number and software version Press DONE to return to the Standby screen R E C O R D T I M E Allows the operator to select a specific recording period or to choose manual start stop To enter the amount of time you wish to record use the NEXT key to select the desir...

Страница 23: ...ssed calibration gas with a nitrogen balance It is highly recommended that a range of calibration gas near the full scale value of the instrument be utilized For example if the instrument is rated for a 0 50 measurement the calibration gas range for spanning the instrument should be between 40 and 50 After sufficient warm up turn the gas selector to ZERO on the conditioning unit and allow the zero...

Страница 24: ...AL GAS to ZERO GAS and then press DONE Once this is completed the analyzer will perform a final zero and then come back into STANBY mode or CONDITION At this time you may turn off the ZERO GAS and put the analyzer back on line by selecting SAMPLE GAS on your gas selector R E C O M M E N D E D C A L I B R A T I O N S C H E D U L E Calibration schedules are typically set forth by the state complianc...

Страница 25: ...nd 30 Once you have set the interval press CONTINUE Set the start date and time for the first auto zero The date is displayed as day month year Use the UP and DOWN buttons to select the day When the day is selected press MONTH to select the month Select the month by pressing the UP and DOWN buttons When done select YEAR to edit the year ...

Страница 26: ... twenty four hour clock as hh mm ss Select the hour by pressing the UP and DOWN buttons When done select MINUTE to set the minutes Select the minute by pressing the up and down buttons When done select Seconds to set the seconds Select the seconds by pressing the up and down buttons When done select Continue to continue ...

Страница 27: ... for the auto calibrating feature to run This must be a value between 1 and 30 It must be a multiple of the number of days between auto zeroing The calibration will immediately follow the auto zero A value of zero will disable auto calibration Select Done when complete ...

Страница 28: ...27 ...

Страница 29: ...put valve and auto drain b A pressure regulator with safety relief valve c A second particulate filter and liquids trap with manual drain d A gas selector valve for zero span and sample gases e A flow indicator with valve for adjusting the sample flow f All stainless steel construction The plate on which the sample conditioner components are mounted provides space and mounting hardware for install...

Страница 30: ... the sample conditioner is designed to remove particulates and condensed vapors from the sample gas The pressure regulator reduces the sample pressure to approximately 8 psig while the pressure relief valve limits the pressure to the flow meter to approximately 15 psig in case of regulator failure The flow meter monitors and is used to adjust the flow of gas to the analyzer to approximately 1 scfh...

Страница 31: ... SPAN position and adjust the pressure regulator on the span gas bottle for an indicated flow of 1 scfh Return the selector valve to the ZERO or OFF position The system is now ready to operate O P E R A T I O N Follow the CALIBRATION AND OPERATION instructions provided earlier in this manual Use the gas selector valve to select the various gases Open the drain valve located below the larger filter...

Страница 32: ...lter holder to free the filter bowl Remove the nut attached to the bottom and remove the filter bowl and element If supplied clean the automatic drain orifice as above Replace the filter element with a new one Infrared Industries part no 273 943 and reinstall the filter bowl in the reverse order as above ...

Страница 33: ...ser should determine the suitability of the product for his or her intended use and the user assumes all risks and liabilities whatsoever in connection therewith W A R R A N T Y E X C L U S I O N S THIS WARRANTY AND THE SELLER AND OR MANUFACTURER S OBLIGATION HEREUNDER IS IN LIEU OF ALL OTHER WARRANTIES EXPRESSED OR IMPLIED INCLUDING WITHOUT LIMITATION THE WARRANTIES OF MERCHANTABILITY AND FITNESS...

Страница 34: ...TO PROPERTY THE LIABILITY OF THE SELLER AND OR MANUFACTURER ON THIS WARRANTY IS LIMITED TO ACCEPTING RETURN OF THE PORTABLE ANALYZER REFUNDING ANY AMOUNT PAID THEREON AND CANCELING ANY BALANCE STILL OWING ON THE EQUIPMENT THIS REMEDY IS EXCLUSIVE REPAIR OR REPLACEMENT PROCEDURE ...

Страница 35: ...bove Package the instrument carefully and securely Contact Infrared Industries for an RMA number which is your authorization to send the unit Note it carefully on the address label Then proceed to ship the complete instrument with freight prepaid to the above address Please include a written description of any observations of the malfunction Repaired instruments shall remain subject to the one yea...

Страница 36: ...35 ...

Страница 37: ...________________________________________________________ SPECIAL DATA _____________________________________________________________________________________ ________________________________________________________________________________________________________ ORIGINAL PURCHASER ___________________________________________________________________________ ____________________________________________...